Running ASTM Method D4815 for Determining Oxygenates in Gasoline on the Agilent 6820 GC Application Note

Running ASTM Method D4815 for Determining Oxygenates in Gasoline on the Agilent 6820 GC Application Note

Application Note 6 pages · 474 KB · EN

The document details a method for determining oxygenates, including methyl tert-butylether (MTBE), ethyl tert-butylether (ETBE), and various alcohols, in unleaded gasoline. The procedure utilizes the Agilent 6820 Gas Chromatography (GC) system equipped with a flame ionization detector (FID), a 10-port switch valve, and two specific columns: a polar TCEP micropacked column for prefractionation and a nonpolar HP-1 capillary column for separation.

The application outlines the experimental configuration, including hardware components like the auxiliary gas pressure regulator and software used (Cerity NDS). It provides specific instrument conditions such as oven programs, flow rates, and temperatures. The text also explains the method description involving valve switching to separate lighter hydrocarbons from oxygenates, presents calibration procedures with standards, and discusses results regarding repeatability which exceeds ASTM D4815 specifications.

Read online

Agilent D4815 specifications

System TypeAgilent 6820 GC system
Prefraction ColumnMicropacked column packed with polar 1,2,3-tris-2-cyanoethoxypropane (TECP)
Analytic ColumnNonpolar HP-1 capillary column
Internal Standard1,2-dimethoxyethane (DME)
Inlet Temperature200 °C
Detector Temperature250 °C

Source: Running ASTM Method D4815 for Determining Oxygenates in Gasoline on the Agilent 6820 GC Application Note

All Agilent D4815 documents by type

User Manual 2

Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.

Application Note 1

How to use the product for one specific task, with worked examples.

Questions answered in these manuals

What compounds are determined by this method?

The specific compounds determined include methyl tert-butylether (MTBE), ethyl tert-butylether (ETBE), tert-amylmethylether (TAME), diisopropylether (DIPE), and various alcohols such as methanol, ethanol, and propanol.

From: Running ASTM Method D4815 for Determining Oxygenates in Gasoline on the Agilent 6820 GC Application Note

What is the recommended backflush time for analyzing DIPE?

For this system, a backflush time of 0.22 minutes is used for the analysis of DIPE to ensure hydrocarbons are completely eluted while ethers are retained.

From: Running ASTM Method D4815 for Determining Oxygenates in Gasoline on the Agilent 6820 GC Application Note