Agilent TC-C18(2) Manuals

ManualLib has 8 documents for the Agilent TC-C18(2) (Other Electronics): 2 product leaflets, 5 application notes. Every file can be read online or downloaded as a PDF for free.

Product Leaflet 2Application Note 5

Key facts

Brand
Agilent
Model
TC-C18(2)
Documents
8
Languages
ES · EN
Last added
2012-07-09

Start here

Agilent TC-C18(2) and HC-C18(2) Reverse Phase LC Columns Brochure

Agilent TC-C18(2) and HC-C18(2) Reverse Phase LC Columns Brochure

Product Leaflet 8 pages · 889 KB · ES

This document introduces the new Agilent TC-C18(2) and HC-C18(2) reverse-phase HPLC columns, highlighting their superior peak shape, reproducibility, and reliability for routine laboratory analysis. Developed to meet high-quality standards, these columns are designed for a wide range of sample types and mobile phases. This manual is written in Spanish.

Read online

Agilent TC-C18(2) specifications

Operating Pressure5000 psi (340 bar)
Column Dimensions4.6 mm id x 12.5 mm
Void Volume< 150 µL
Maximum Temperature60 °C

Source: Agilent HC-C18(2) TC-C18(2) High Performance Reliance Cartridge Guard-Columns Manual

All Agilent TC-C18(2) documents by type

Questions answered in these manuals

What is the stated particle size for both the TC-C18(2) and HC-C18(2) columns?

Both the TC-C18(2) and HC-C18(2) columns have a stated particle size of 5 um.

From: Agilent TC-C18(2) and HC-C18(2) Reverse Phase LC Columns Brochure

What is the specified pH range for using the HC-C18(2) column?

The HC-C18(2) column is stable over a stated pH range of 2-9.

From: Agilent TC-C18(2) and HC-C18(2) Reverse Phase LC Columns Brochure

How often should the guard cartridge be replaced?

Replace the guard cartridge when system pressure has increased by 10% from normal or after every 100 to 200 injections.

From: Agilent HC-C18(2) TC-C18(2) High Performance Reliance Cartridge Guard-Columns Manual

What is the recommended procedure for flushing a new guard column?

Flush a minimum of 3 mL of methanol through the assembly, then flush with 10 mL of mobile phase.

From: Agilent HC-C18(2) TC-C18(2) High Performance Reliance Cartridge Guard-Columns Manual

Which software controls the instrument?

The instrument is controlled by Agilent EZChrom Elite Compact Compliance software.

From: Agilent In-process control for Aripiprazole manufacturing with the Agilent 1120 Compact LC Agilent TC C18(2) columns Manual

What is the total run time including equilibration?

The total run time, including equilibration and re-equilibration, can be limited to 14 minutes.

From: Agilent In-process control for Aripiprazole manufacturing with the Agilent 1120 Compact LC Agilent TC C18(2) columns Manual