Agilent TC-C18(2) and HC-C18(2) Reverse Phase LC Columns Brochure
This document introduces the new Agilent TC-C18(2) and HC-C18(2) reverse-phase HPLC columns, highlighting their superior peak shape, reproducibility, and reliability for routine laboratory analysis. Developed to meet high-quality standards, these columns are designed for a wide range of sample types and mobile phases. This manual is written in Spanish.
Read onlineAgilent TC-C18(2) specifications
| Operating Pressure | 5000 psi (340 bar) |
|---|---|
| Column Dimensions | 4.6 mm id x 12.5 mm |
| Void Volume | < 150 µL |
| Maximum Temperature | 60 °C |
Source: Agilent HC-C18(2) TC-C18(2) High Performance Reliance Cartridge Guard-Columns Manual
All Agilent TC-C18(2) documents by type
How to use the product for one specific task, with worked examples.
- Agilent In-process control for Aripiprazole manufacturing with the Agilent 1120 Compact LC Agilent TC C18(2) columns Manual 4 p
- Agilent Detector linearity testing for Aripiprazole quality control with the Agilent 1120 Compact LC Agilent TC C18(2) columns Manual 4 p
- Agilent System suitability testing for Aripiprazole quality control with the Agilent 1120 Compact LC Agilent TC C18(2) columns Manual 4 p
- Agilent TC-C18(2) Column Application Note: Separation of Seven Synthetic Food Colors 8 p
- Agilent Injector linearity testing for Tramadol quality control with the Agilent 1120 Compact LC Agilent TC C18(2) columns Application Note 8 p
Questions answered in these manuals
What is the stated particle size for both the TC-C18(2) and HC-C18(2) columns?
Both the TC-C18(2) and HC-C18(2) columns have a stated particle size of 5 um.
From: Agilent TC-C18(2) and HC-C18(2) Reverse Phase LC Columns Brochure
What is the specified pH range for using the HC-C18(2) column?
The HC-C18(2) column is stable over a stated pH range of 2-9.
From: Agilent TC-C18(2) and HC-C18(2) Reverse Phase LC Columns Brochure
How often should the guard cartridge be replaced?
Replace the guard cartridge when system pressure has increased by 10% from normal or after every 100 to 200 injections.
From: Agilent HC-C18(2) TC-C18(2) High Performance Reliance Cartridge Guard-Columns Manual
What is the recommended procedure for flushing a new guard column?
Flush a minimum of 3 mL of methanol through the assembly, then flush with 10 mL of mobile phase.
From: Agilent HC-C18(2) TC-C18(2) High Performance Reliance Cartridge Guard-Columns Manual
Which software controls the instrument?
The instrument is controlled by Agilent EZChrom Elite Compact Compliance software.
What is the total run time including equilibration?
The total run time, including equilibration and re-equilibration, can be limited to 14 minutes.