Agilent 6540 Accurate-Mass Q-TOF Technical Overview
The document presents a technical overview demonstrating how the Agilent 6540 Accurate-Mass Q-TOF LC/MS System can perform metabolic stability, profiling, and identification in a single experiment. This approach maximizes information obtained from one analysis compared to using separate methods for quantitation and identification.
Read onlineAgilent 6540 Accurate-Mass Q-TOF specifications
| Drying Gas Temperature | 350°C |
|---|---|
| Sheath Gas Flow | 12 L/min |
| Capillary Voltage | 4000 V |
| Mass Range (MS and MS/MS) | 100-1000 m/z |
| Acquisition Rate (Combined Mode) | 5 spectra/sec |
| Collision Energy | 28 eV |
All Agilent 6540 Accurate-Mass Q-TOF documents by type
How to use the product for one specific task, with worked examples.
Questions answered in these manuals
What modes does the Agilent 6540 operate in for this study?
The system was operated in an initial MS-only mode at 5 Hz acquisition rate and an automatic MS/MS mode with expected metabolites on the inclusion list.
How does the Agilent 6540 compare to the Triple Quadrupole system?
The quantitative data obtained using the 6540 QTOF showed good correlation with analysis performed on the Agilent 6460 Triple Quadrupole LC/MS System.
What is the mass error in MS mode for the example compound?
The mass error was only 0.33 ppm.
What is the MS/MS mass accuracy above m/z 100?
The MS/MS mass accuracy is better than 2 ppm.