Agilent 720-ES 730-ES Series SmartRinse Maximizing Throughput Application Note Manual

Agilent 720-ES 730-ES Series SmartRinse Maximizing Throughput Application Note Manual

Application Note 4 pages · 111 KB · EN

SmartRinse is a feature within the Agilent ICP Expert II software designed to speed up sample wash-out in 700 Series ICP-OES instruments. It reduces carry-over and increases productivity by optimizing the rinse time between samples based on the specific element concentrations being measured, rather than using a fixed, worst-case time for all samples.

The feature works by measuring the baseline signal of selected element emission lines and calculating a threshold intensity. During the rinse phase, the system continuously monitors the signal intensity. Rinsing stops automatically once the signals for all selected elements fall below their calculated thresholds. The sensitivity of this process can be adjusted by the user via an SD factor setting to accommodate different analytical requirements.

Read online

Agilent 720-ES specifications

Power1.2 kW
Plasma gas flow15 L/min
Auxiliary gas flow1.5 L/min
Nebulizer flow0.65 L/min
Pump speed12 rpm
Sample uptake rate1.0 mL/min
Replicate read time30s
Sample uptake delay time80s
Rinse time120s

Source: Agilent 720-ES Axial ICP-OES Mercury Determination Application Note

All Agilent 720-ES documents by type

Questions answered in these manuals

How does SmartRinse determine when to stop rinsing?

SmartRinse stops rinsing when the measured signal intensity of all nominated emission lines falls below a pre-determined threshold intensity calculated from baseline measurements.

From: Agilent 720-ES 730-ES Series SmartRinse Maximizing Throughput Application Note Manual

Can the sensitivity of SmartRinse be adjusted?

Yes, the sensitivity can be adjusted by changing the SD factor for each element in the ICP Expert II software to set a higher or lower threshold intensity.

From: Agilent 720-ES 730-ES Series SmartRinse Maximizing Throughput Application Note Manual

How should the torch be cleaned after analysis?

Soak the torch in concentrated HCl to dissolve the white SnO2 deposit, then rinse with de-ionized water and dry.

From: Agilent 720-ES Axial ICP-OES Mercury Determination Application Note

What is the recommended procedure for cleaning white powder deposits from the end of the torch after analysis?

To remove the deposit, soak the torch directly in concentrated HCl to dissolve it, then rinse with de-ionized water and allow it to dry.

From: Agilent 720-ES Axial ICP-OES Mercury Determination Application Note