Agilent Cation Solution Kit for the Agilent 7100 CE System Technical Note
This document is a technical note describing the use of the Agilent Cation Solution Kit with the Agilent 7100 Capillary Electrophoresis System. The kit contains necessary supplies, including buffers and reagents, ready-to-use for the separation of cations in indirect UV detection mode. It is also suitable for separating low-molecular weight amines.
The note details an experimental method using a DAD detector with a 56-cm effective length capillary. It provides specific settings for the instrument setup, including voltage (30 KV), current (300 pA), and injection parameters. The document also includes data on migration time reproducibility and tap water analysis results.
Agilent 7100 CE specifications
| Part Number | 5064-8206 |
|---|---|
| Voltage | 30 KV |
| Current | 300 pA |
| Power | 6w |
| Stoptime | 8:00 min |
Source: Agilent Cation Solution Kit for the Agilent 7100 CE System Technical Note
All Agilent 7100 CE documents by type
How to use the product for one specific task, with worked examples.
Questions answered in these manuals
What is the part number of the Cation Solutions Kit?
The part number for the Cation Solutions Kit is 5064-8206.
From: Agilent Cation Solution Kit for the Agilent 7100 CE System Technical Note
What cations are included in the test sample?
The Cation Test Sample includes ammonium, potassium, sodium, calcium, and magnesium at 100 ppm each.
From: Agilent Cation Solution Kit for the Agilent 7100 CE System Technical Note
What is the minimum concentration of a spiked protein that can be detected using these gels?
It is possible to detect the spiked protein down to a level of 0.1 %.
What temperature is maintained for the capillary during runs with Advanced Analytical gels?
The capillary temperature was kept at 25 °C.