Agilent 380-ELSD Manuals

ManualLib has 4 documents for the Agilent 380-ELSD (Other Electronics): 1 product leaflet, 1 datasheet, 2 application notes. Every file can be read online or downloaded as a PDF for free.

Product Leaflet 1Datasheet 1Application Note 2

Key facts

Brand
Agilent
Model
380-ELSD
Documents
4
Languages
EN
Last added
2012-07-05

Start here

Agilent 380-ELSD Optimization of ELSD Conditions using PEG 400 Technical Overview

Agilent 380-ELSD Optimization of ELSD Conditions using PEG 400 Technical Overview

Product Leaflet 2 pages · 561 KB · EN

This document provides a technical overview of the Agilent 380-ELSD, highlighting its suitability for analyzing non-UV absorbing compounds such as polyethylene glycols (PEGs). It explains that detection sensitivity can be maximized by tuning three main control variables: nebulizer temperature, evaporator temperature, and gas flow rate.

Read online

Agilent 380-ELSD specifications

Operating Temperatureup to 120 °C
Reproducibilitybelow 2%
Detection Rangelow nanogram levels

Source: Agilent 380-ELSD Evaporative Light Scattering Detector Data Sheet

All Agilent 380-ELSD documents by type

Datasheet 1

Ratings, dimensions and ordering codes, for selection and design work.

Questions answered in these manuals

What are the three main control variables for the Agilent 380-ELSD?

The three main control variables are nebulizer temperature, evaporator temperature, and gas flow rate.

From: Agilent 380-ELSD Optimization of ELSD Conditions using PEG 400 Technical Overview

What effect does a higher evaporator temperature have on the analysis?

A higher evaporator temperature results in a visible increase in signal and a reduction in baseline noise.

From: Agilent 380-ELSD Optimization of ELSD Conditions using PEG 400 Technical Overview

What compounds is the Agilent 380-ELSD ideal for?

It is ideal for detecting non-volatile compounds, where temperatures of 100 °C or above are required.

From: Agilent 380-ELSD Evaporative Light Scattering Detector Data Sheet

Does the Agilent 380-ELSD require an analog to digital converter?

No, control and digital data collection come as standard for multi-vendor platforms, so there is no need for an analog to digital converter.

From: Agilent 380-ELSD Evaporative Light Scattering Detector Data Sheet

What is the Agilent 380-ELSD ideal for analyzing?

The system is an ideal solution for the characterization of xylooligosaccharides and other water-soluble sugar oligomers.

From: Agilent 380-ELSD Application Note for Xylooligosaccharide Characterization

What are the recommended operating conditions for flow rate and injection volume?

The procedure specifies a flow rate of 1.0 mL/min and an injection volume of 20 uL.

From: Agilent 380-ELSD Application Note for Xylooligosaccharide Characterization