Agilent J&W DB-5ms Ultra Inert Column Application Note
This document presents a comparative study on the chromatographic behavior of Agilent J&W DB-5ms Ultra Inert columns versus similar columns from four other manufacturers. The analysis focuses on the inertness required for analyzing very active samples, specifically basic compounds that are prone to tailing due to acidic sites in non-inert columns.
The study utilized an Agilent 6890N GC system to test a mixture of six basic compounds, chlorophenols, organophosphorus pesticides, and aromatic positional isomers. Results demonstrated that the DB-5ms Ultra Inert column provided superior peak shapes with significantly reduced tailing factors compared to the competitor columns, ensuring more accurate and reliable analytical results.
Agilent W DB-5ms Ultra Inert specifications
| Column Length | 15 m |
|---|---|
| Column Inner Diameter | 0.25 mm |
| Column Film Thickness | 0.25 µm |
| Column Part Number | 122-5512UI |
| Carrier Gas Flow Rate | 72 cm/s |
| Inlet Temperature | 325 °C |
| Oven Temperature Range | 150 to 325 °C |
| MSD Source Temperature | 300 °C |
Source: Agilent J&W DB-5ms Ultra Inert Capillary GC Column PBDE Analysis Application Note
All Agilent W DB-5ms Ultra Inert documents by type
How to use the product for one specific task, with worked examples.
- Agilent J&W DB-5ms Ultra Inert Column Application Note 8 p
- Agilent J&W DB-5ms Ultra Inert Capillary GC Column PBDE Analysis Application Note 6 p
- Agilent Trace-Level Analysis of Melamine in Milk Products on Agilent 7890A/5975C GC/MSD Using a New Agilent J W DB-5ms Ultra Inert Column SampliQ SCX Cartridges Manual 6 p
- Agilent J&W DB-5ms Ultra Inert Capillary GC Column Application Note for Organophosphorus Pesticides Analysis 8 p
- Agilent J&W DB-5ms Ultra Inert Capillary GC Column Application Brief for PAH Analysis 2 p
- Agilent J&W DB-5ms Ultra Inert GC Column PBDE Analysis Application Note 2 p
Questions answered in these manuals
What specific compounds were used to test the column's performance with basic analytes?
The study analyzed a mixture containing Triethylamine, Pyridine, 4-Methylpyridine, N.N-Dimethylacetamide, 2,4-Dimethylaniline, and Dicyclohexylamine.
From: Agilent J&W DB-5ms Ultra Inert Column Application Note
What is the length of the Agilent J&W DB-5ms Ultra Inert column used in this study?
The column dimensions are 30 m x 0.25 mm x 0.25 um.
From: Agilent J&W DB-5ms Ultra Inert Column Application Note
Which Agilent column is recommended for challenging PBDE analysis?
The document recommends the Agilent J&W DB-5ms Ultra Inert capillary GC column.
From: Agilent J&W DB-5ms Ultra Inert Capillary GC Column PBDE Analysis Application Note
What are the concentration ranges used for BDE calibration?
The study used a range of 0.5 to 250 ng/mL for most BDEs and 2.5 to 1,000 ng/mL for BDE-209.
From: Agilent J&W DB-5ms Ultra Inert Capillary GC Column PBDE Analysis Application Note
What is the lowest concentration of melamine this method can detect in milk?
The limit of detection for melamine in the milk matrix is 0.0025 mg/kg.
What is the linear concentration range for this melamine analysis method?
The method shows good linearity in the range of 0.025 to 2.000 pg/mL.