Agilent PL-GPC 50 Plus Manuals

ManualLib has 10 documents for the Agilent PL-GPC 50 Plus: 10 application notes. Every file can be read online or downloaded as a PDF for free.

Application Note 10

Key facts

Brand
Agilent
Model
PL-GPC 50 Plus
Documents
10
Languages
EN
Last added
2012-06-12

Start here

Agilent PL-GPC 50 Plus Poly(styrene-co-butadiene) GPC Application Note

Agilent PL-GPC 50 Plus Poly(styrene-co-butadiene) GPC Application Note

Application Note 3 pages · 530 KB · EN

This document details the analysis of styrene/butadiene rubber copolymers using gel permeation chromatography (GPC) with viscometry. It explains that while light scattering is unsuitable for these copolymers due to non-uniform chemistry, the Universal Calibration technique using a viscometer provides accurate molecular weight distributions independent of column standards.

The analysis was performed on an Agilent PL-GPC 50 Plus integrated system equipped with a differential refractive index detector, an Agilent PL-BV AQ0RT viscometer, and PLgel 5 pm MIXED-C columns. The results demonstrate the system's ability to resolve molecular weight distributions and determine material structure for elastomeric materials.

Read online

Agilent PL-GPC 50 Plus specifications

Flow Rate1 mL/min
Temperature40°C
Column Dimensions300 x 7.5 mm
Operating Temperature Rangeambient to 50 °C

Source: Agilent PL-GPC 50 Plus Poly(styrene-co-butadiene) GPC Application Note

All Agilent PL-GPC 50 Plus documents by type

Questions answered in these manuals

Why is viscometry used instead of light scattering for these samples?

Light scattering cannot be used because the samples are copolymers with non-uniform chemistry, which leads to errors in calculations.

From: Agilent PL-GPC 50 Plus Poly(styrene-co-butadiene) GPC Application Note

What advantage does the Universal Calibration technique offer?

It gives molecular weight that is independent of the standards used in the column calibration.

From: Agilent PL-GPC 50 Plus Poly(styrene-co-butadiene) GPC Application Note

What instrument was used to analyze the cornflour samples?

The samples were analyzed by an Agilent PL-GPC 50 Plus fitted with an Agilent PL-BV 400RT viscometer.

From: Agilent PL-GPC 50 Plus Application Note: Analysis of Cornflour by GPC with Viscometry

How was the structure of the cornflour samples compared?

The structure was compared using the Mark-Houwink plot of log (intrinsic viscosity) as a function of log (molecular weight).

From: Agilent PL-GPC 50 Plus Application Note: Analysis of Cornflour by GPC with Viscometry

Does the PL-GPC 50 Plus require column calibration for triple detection analysis?

No, using triple detection allows the sample to be analyzed without the need to perform a column calibration.

From: Agilent PL-GPC 50 Plus Aqueous SEC Triple Detection Application Note

What is the operating temperature range of the PL-GPC 50 Plus?

The system is designed for operation from ambient to 50°C.

From: Agilent PL-GPC 50 Plus Aqueous SEC Triple Detection Application Note