Agilent 7890A GC and 5975C MSD On-Site Training Manual
This on-site training manual helps users operate the Agilent 7890A gas chromatograph and 5975C mass spectrometer. It covers instrument configuration, startup and shutdown, column aging, vacuum status, autotune and standard spectra tune, and GC configuration editing.
The manual also describes ALS, module, and column parameter settings, chemical ionization operation with methane reagent gas, and venting the system. It is intended for the 1701EA series workstation and includes Chinese and English interface references. This manual is written in Chinese.
Agilent 7890A-5975C specifications
| GC model | 7890A |
|---|---|
| MSD model | 5975C |
| Workstation | 1701EA series |
| Carrier gas | Helium (He) |
| Regulator pressure | 0.5 MPa |
| Operating system | Windows XP (SP2) |
| Turbomump speed indicator | 100 Percent |
| Diffusion pump pressure | About 100 mTorr |
| Minimum time before tuning | At least 2 hours |
| Autotune duration | About 3–5 minutes |
| Methane flow ratio | m/z 28/27 ratio of 1.5–5.0 |
| CI tuning standard | PFDTD, molecular formula C11HF23O3, molecular weight 618.12 |
Source: Agilent 7890A GC and 5975C MSD On-Site Training Manual
All Agilent 7890A-5975C documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
How to use the product for one specific task, with worked examples.
- Agilent Trace-Level Analysis of Melamine in Milk Products on Agilent 7890A/5975C GC/MSD Using a New Agilent J W DB-5ms Ultra Inert Column SampliQ SCX Cartridges Manual 6 p
- Agilent Rapid Screening of Melamine Cyanuric Acid in Milk Products Using Agilent J W HP-5ms GC Column Agilent 7890A/5975C GC/MSD with Column Backflushing Manual 8 p
- Agilent 7890A-5975C and 1200SL-6410 Melamine Cyanuric Acid Analysis Application Note 8 p
Questions answered in these manuals
How long should the instrument run before tuning?
Tuning should be performed after the instrument has been on for at least 2 hours. This wait is extended to 4 hours if the instrument has been off for a long time or uses a 5975 diffusion pump configuration.
From: Agilent 7890A GC and 5975C MSD On-Site Training Manual
What is the difference between Autotune and Standard spectra tune?
Autotune provides maximum sensitivity, while Standard spectra tune has slightly lower sensitivity but matches standard library spectra better, making it suitable for library searching and qualitative identification.
From: Agilent 7890A GC and 5975C MSD On-Site Training Manual
What is the lowest concentration of melamine this method can detect in milk?
The limit of detection for melamine in the milk matrix is 0.0025 mg/kg.
What is the linear concentration range for this melamine analysis method?
The method shows good linearity in the range of 0.025 to 2.000 pg/mL.
How much time does the backflushing method save compared to conventional baking?
The method reduces the GC run time from more than 70 minutes to 14.5 minutes.
What is the recovery rate for melamine and cyanuric acid in spiked samples?
The recoveries are greater than 95%, with specific results of 96.1% for cyanuric acid and 95.6% for melamine.