Agilent G6410AA Triple Quadrupole Mass Spectrometer Application Note
This document presents a fast, sensitive, and reproducible technique for confirming the presence of drugs of abuse (DOA) in oral fluids (OF) using the Agilent G6410AA Triple Quadrupole Mass Spectrometer. The method utilizes liquid chromatography/tandem mass spectrometry (LC/MS/MS) with a 1.8-µm, C18 column to achieve high chromatographic resolution with elution times of less than 4.2 minutes for THC and less than 1.5 minutes for other analyzed drugs. The technique meets the sensitivity requirements set by the United States Substance Abuse and Mental Health Services Administration (SAMHSA) for workplace testing.
Read onlineAgilent G6410AA specifications
| Column temperature | 40°C |
|---|---|
| Flow rate | 0.5 mL/min |
| Elution time (THC) | Less than 4.2 minutes |
| Elution time (Cocaine, Amphetamine, Methamphetamine, MDMA) | Less than 1.5 minutes |
Source: Agilent G6410AA Triple Quadrupole Mass Spectrometer Application Note
All Agilent G6410AA documents by type
How to use the product for one specific task, with worked examples.
- Agilent G6410AA Triple Quadrupole Mass Spectrometer Application Note 8 p
- Agilent G6410AA LC/MS/MS Aflatoxin Determination Application Note 8 p
- Agilent G6410AA LC/MS Application Note: Chloramphenicol Analysis in Food 8 p
- Agilent G6410AA Triple Quadrupole Mass Spectrometer Pesticide Determination Application Note 12 p
Questions answered in these manuals
What is the elution time for THC in this method?
The elution time for THC is less than 4.2 minutes.
From: Agilent G6410AA Triple Quadrupole Mass Spectrometer Application Note
What is the elution time for cocaine, amphetamine, methamphetamine, and MDMA?
The elution time for these drugs is less than 1.5 minutes.
From: Agilent G6410AA Triple Quadrupole Mass Spectrometer Application Note
Which aflatoxins are determined in this method?
The method determines aflatoxin G1, G2, B1, and B2 residues in cereals.
From: Agilent G6410AA LC/MS/MS Aflatoxin Determination Application Note
What is the limit of detection for the aflatoxins?
The limits of detection for all aflatoxins were less than 1 ng/mL in cereals.
From: Agilent G6410AA LC/MS/MS Aflatoxin Determination Application Note
What is the minimum detection limit for chloramphenicol in honey, shrimp, and chicken using this method?
The method provides a limit of detection at 10 ppt in the sample matrix.
From: Agilent G6410AA LC/MS Application Note: Chloramphenicol Analysis in Food
What is the maximum flow rate used for the HPLC system in this analysis?
The flow rate for the HPLC system is set to 0.4 mL/min.
From: Agilent G6410AA LC/MS Application Note: Chloramphenicol Analysis in Food