Running ASTM Method D4815 for Determining Oxygenates in Gasoline on the Agilent 6820 GC Application Note

The document describes an analytical method for the determination of ethers and alcohols in gasolines using the Agilent 6820 gas chromatography system. This method utilizes a 10-port switch valve and two columns. The specific compounds determined include methyl tert-butylether (MTBE), ethyl tert-butylether (ETBE), tert-amylmethylether (TAME), diisopropylether (DIPE), methanol, ethanol, isopropanol, n-propanol, isobutanol, tert-butanol, sec-butanol, n-butanol, and tert-pentanol. 1,2-dimethoxyethane (DME) is used as the internal standard for quantitative analysis. The Agilent 6820 GC is used to run ASTM Method D4815 for determining oxygenates in gasoline.


Running ASTM Method D4815 for Determining Oxygenates in Gasoline on the Agilent 6820 GC Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
474 KB
Published
02 July, 2012
Updated
29 September, 2023
Models
D4815, 6820
MD5 checksum
6691F59CDCA253C8DB13FFDAD10BE814

About this manual

The document details a method for determining oxygenates, including methyl tert-butylether (MTBE), ethyl tert-butylether (ETBE), and various alcohols, in unleaded gasoline. The procedure utilizes the Agilent 6820 Gas Chromatography (GC) system equipped with a flame ionization detector (FID), a 10-port switch valve, and two specific columns: a polar TCEP micropacked column for prefractionation and a nonpolar HP-1 capillary column for separation.

The application outlines the experimental configuration, including hardware components like the auxiliary gas pressure regulator and software used (Cerity NDS). It provides specific instrument conditions such as oven programs, flow rates, and temperatures. The text also explains the method description involving valve switching to separate lighter hydrocarbons from oxygenates, presents calibration procedures with standards, and discusses results regarding repeatability which exceeds ASTM D4815 specifications.

What's inside

  • Running ASTM Method D4815
  • Determining Oxygenates in Gasoline
  • Agilent 6820 GC System Configuration
  • Experimental Setup and Hardware
  • Method Description and Valve Switching
  • Backflush Time Setting
  • Instrument Conditions
  • Calibration and Standardization
  • Results and Repeatability

Specifications

System TypeAgilent 6820 GC system
Prefraction ColumnMicropacked column packed with polar 1,2,3-tris-2-cyanoethoxypropane (TECP)
Analytic ColumnNonpolar HP-1 capillary column
Internal Standard1,2-dimethoxyethane (DME)
Inlet Temperature200 °C
Detector Temperature250 °C

Frequently asked questions

What compounds are determined by this method?

The specific compounds determined include methyl tert-butylether (MTBE), ethyl tert-butylether (ETBE), tert-amylmethylether (TAME), diisopropylether (DIPE), and various alcohols such as methanol, ethanol, and propanol.

What is the recommended backflush time for analyzing DIPE?

For this system, a backflush time of 0.22 minutes is used for the analysis of DIPE to ensure hydrocarbons are completely eluted while ethers are retained.

Does the method meet ASTM D4815 specifications?

Yes, the experimental data demonstrates that the results exceed the ASTM D4815 calibration and repeatability specifications.