Agilent Improved Performance for the Analysis of Aromatics in Gasoline by ASTM Method D5769 Using the Agilent 5973 inert GC/MS System Manual
The document details the use of the Agilent 5973 inert gas chromatography/mass spectrometry system to measure high concentrations of aromatic compounds in gasoline according to ASTM Method D5769. It highlights that the optional extended linearity ion source kit provides improved linear calibrations, making the method easier and faster to set up with more reliable results compared to standard high-sensitivity sources.
Read onlineAgilent D-5769 specifications
| Carrier gas | Helium |
|---|---|
| Inlet temperature | 250 °C |
| Oven temperature program | 60 °C (0 min) to 250 °C (0 min) |
| Split ratio | 250:1 |
| Electron energy | 70 eV |
| Scan range | 45 to 300 Daltons |
| Analysis range | 0.1 to 42 weight percent |
All Agilent D-5769 documents by type
How to use the product for one specific task, with worked examples.
Questions answered in these manuals
What ASTM method is used for analyzing aromatics in gasoline?
The document uses American Society of Testing and Materials Method D5769.
Does the extended linearity source improve calibration?
Yes, it provides greater linear dynamic range, resulting in a coefficient of determination (r²) of 0.997 for toluene compared to 0.911 for the standard source.
What is the recommended split ratio for the GC injection port?
The recommended split ratio is 1100 to 1.
From: Agilent Technologies 6890/5973 GC/MSD Application Note ASTM D-5769 Gasoline Analysis Enhancements
What is the total analysis time for the gasoline method?
The analysis time is 20 minutes.
From: Agilent Technologies 6890/5973 GC/MSD Application Note ASTM D-5769 Gasoline Analysis Enhancements