Agilent Improved Performance for the Analysis of Aromatics in Gasoline by ASTM Method D5769 Using the Agilent 5973 inert GC/MS System Manual

This application describes the improved performance for the analysis of aromatics in gasoline by ASTM Method D5769 using the Agilent 5973 inert gas chromatography/mass spectrometry (GC/MS) system.


Agilent Improved Performance for the Analysis of Aromatics in Gasoline by ASTM Method D5769 Using the Agilent 5973 inert GC/MS System Manual - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
782 KB
Published
02 July, 2012
Updated
30 September, 2023
Model
D-5769
MD5 checksum
C88F4E7BE5B5B691055B86A99E08C83D

About this manual

The document details the use of the Agilent 5973 inert gas chromatography/mass spectrometry system to measure high concentrations of aromatic compounds in gasoline according to ASTM Method D5769. It highlights that the optional extended linearity ion source kit provides improved linear calibrations, making the method easier and faster to set up with more reliable results compared to standard high-sensitivity sources.

What's inside

  • Abstract
  • Introduction
  • Experimental
  • Results and Discussion
  • Conclusion
  • References

Specifications

Carrier gasHelium
Inlet temperature250 °C
Oven temperature program60 °C (0 min) to 250 °C (0 min)
Split ratio250:1
Electron energy70 eV
Scan range45 to 300 Daltons
Analysis range0.1 to 42 weight percent

Frequently asked questions

What ASTM method is used for analyzing aromatics in gasoline?

The document uses American Society of Testing and Materials Method D5769.

Does the extended linearity source improve calibration?

Yes, it provides greater linear dynamic range, resulting in a coefficient of determination (r²) of 0.997 for toluene compared to 0.911 for the standard source.