Agilent 5975C MSD Manuals

ManualLib has 2 documents for the Agilent 5975C MSD: 2 application notes. Every file can be read online or downloaded as a PDF for free.

Application Note 2

Key facts

Brand
Agilent
Model
5975C MSD
Documents
2
Languages
EN
Last added
2012-07-02

Start here

Capillary Flow Technology for GC/MS: Efficacy of the Simple Tee Configuration for Robust Analysis Using Rapid Backflushi

Capillary Flow Technology for GC/MS: Efficacy of the Simple Tee Configuration for Robust Analysis Using Rapid Backflushi

Application Note 4 pages · 156 KB · EN

This document describes an application of Capillary Flow Technology (CFT) for Gas Chromatography-Mass Spectrometry (GC/MS). It illustrates how a pressure-controlled tee (PCT) arrangement, utilizing a purged union and electronic pressure control devices, can eliminate late-eluting matrix components without venting the mass spectrometer. The method uses rapid backflushing to remove high-boiling contaminants that typically foul the ion source.

Read online

Agilent 5975C MSD specifications

Sample Volume10-mL water samples
Vial Type20-mL headspace vial
Internal Standard Concentration800 ng/L (800 ppt)
Target Analytes60 target analytes
Calibration Levels45 to 1,250 ng/L
Column TypeDB-624, 20 m x 0.18 mm
Carrier GasHe, constant pressure (95 kPa)
Run Time25-min run

Source: Agilent 7890A GC/5975C MSD VOC Analysis Application Note

All Agilent 5975C MSD documents by type

Questions answered in these manuals

What is the primary benefit of using a pressure-controlled tee (PCT) arrangement?

It provides rapid backflushing to eliminate late-eluting matrix components, improving run-to-run response robustness and reducing ion source fouling.

From: Capillary Flow Technology for GC/MS: Efficacy of the Simple Tee Configuration for Robust Analysis Using Rapid Backflushi

How does backflushing affect analyte signal stability over multiple injections?

Using PCT with backflushing maintains signal within about 10 percent of the first injection's response, whereas continuous columns without backflushing can lose more than 30 percent.

From: Capillary Flow Technology for GC/MS: Efficacy of the Simple Tee Configuration for Robust Analysis Using Rapid Backflushi

What system was used for the analysis?

The analyses were performed on an Agilent 7890A GC/5975C MSD system.

From: Agilent 7890A GC/5975C MSD VOC Analysis Application Note

What reference water was used for calibration?

Bottled drinking water (Evian) was used for reference water blanks.

From: Agilent 7890A GC/5975C MSD VOC Analysis Application Note