Agilent G6410AA LC/MS/MS Aflatoxin Determination Application Note

A sensitive and selective analytical method for the determination of aflatoxin G1, G2, B1, and B2 residues in cereals using the Agilent G6410AA LC/MS Triple Quadrupole Mass Spectrometer was developed. This method uses simple sample preparation methods followed by LC/MS/MS. The limits of detection for all aflatoxins were less than 1 ng/mL in cereals.


Agilent G6410AA LC/MS/MS Aflatoxin Determination Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
228 KB
Published
12 June, 2012
Updated
28 September, 2023
MD5 checksum
13A7C47322433BF4A7DEBDE54E06D2ED

About this manual

This document details a sensitive and selective analytical method for determining aflatoxin G1, G2, B1, and B2 residues in cereals. The method utilizes the Agilent G6410AA LC/MS Triple Quadrupole Mass Spectrometer following simple sample preparation steps.

The study optimized Liquid Chromatography (LC) and Mass Spectrometry (MS) conditions, including the use of a ZORBAX Extend C18 column and specific MRM transitions. The developed method achieved limits of detection less than 1 ng/mL for all analyzed aflatoxins.

What's inside

  • Abstract
  • Introduction
  • Sample Preparation
  • Standard Preparation
  • Chemicals
  • LC/MS Instrument
  • LC Conditions
  • MS Conditions
  • Results and Discussion
  • Linearity and LODs

Specifications

InstrumentAgilent G6410 Triple Quadrupole Mass Spectrometer
ColumnZORBAX Extend C18, 100 mm x 2.1 mm, 1.8 um
Column temperature40°C
Mobile phase A10mM ammonium acetate in water
Mobile phase BMethanol
Flow rate0.2 mL/min
SourcePositive ESI
Drying gas flow10 L/min
Veap4000 V
Limit of detection (LOD)Less than 1 ng/mL

Frequently asked questions

Which aflatoxins are determined in this method?

The method determines aflatoxin G1, G2, B1, and B2 residues in cereals.

What is the limit of detection for the aflatoxins?

The limits of detection for all aflatoxins were less than 1 ng/mL in cereals.

What column is used for the analysis?

A ZORBAX Extend C18 column with dimensions 100 mm x 2.1 mm and 1.8 um particle size is used.