Key facts
- Brand
- Agilent
- Model
- 6210 TOF
- Documents
- 4
- Languages
- EN
- Last added
- 2012-07-03
Start here

Agilent 6210 TOF LC/MS System Quick Start Guide
This document serves as a roadmap for users to get started with the Agilent 6210 TOF LC/MS System. It provides brief instructions on preparing the instrument, setting up acquisition methods using MassHunter Software, and running interactive samples or worklists.
The guide outlines the initial steps required after hardware and software installation are complete. Key topics include starting the software engine, configuring the instrument status pane to ensure correct LC modules are selected, monitoring real-time plots of chromatograms and spectra, and reviewing acquired data.
Agilent 6210 TOF specifications
| Enrichment column volume | 40-nL |
|---|---|
| Analytical column dimensions | 150 mm x 75 pm |
| Column material | Zorbax 300SB-C18, 5 pm |
| Gradient method duration | 100-minute |
| Analytical flow rate | 300 nL/min |
| Sample loading flow rate | 4 pL/min |
Source: Agilent 1200 Series HPLC-Chip/6210 TOF LC/MS Application Note
All Agilent 6210 TOF documents by type
How to use the product for one specific task, with worked examples.
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
What specific instrument is the MassHunter Profiling software designed for?
The software is specifically designed for use with the Agilent 6210 TOF LC/MS.
From: Agilent MassHunter Profiling Software for the Agilent 6210 TOF LC/MS Manual
What types of applications does the software support besides biomarker discovery?
It supports pharmaceutical impurity analysis, forensic analysis, environmental analysis, and pesticide analysis.
From: Agilent MassHunter Profiling Software for the Agilent 6210 TOF LC/MS Manual
What system is used for the protein profiling approach?
The approach uses an Agilent 1200 Series HPLC-Chip/MS system interfaced to an Agilent 6210 TOF mass spectrometer.
From: Agilent 1200 Series HPLC-Chip/6210 TOF LC/MS Application Note
What software is used for data extraction and differential analysis?
Mass Profiler software is used for feature extraction and differential analysis of the complex data sets.
From: Agilent 1200 Series HPLC-Chip/6210 TOF LC/MS Application Note
What is the maximum temperature applied in the experiment?
A maximum temperature of 80 °C was applied to reduce solvent viscosity.
From: Agilent 1200 Series LC and 6210 TOF MS Application Note
What is the maximum flow rate tolerated by the standard ESI source?
The standard ESI source specifies a maximum flow rate of up to 1 mL/min.