Agilent Zorbax SPE C18 Cartridge Application Brief for Naproxen Analysis

Agilent Zorbax SPE C18 Cartridge Application Brief for Naproxen Analysis

Application Note 2 pages · 35 KB · EN

The document details an analytical procedure for determining levels of naproxen, a pharmaceutical compound used for pain relief, in blood serum. The primary challenge addressed is the interference caused by high protein content in the sample matrix.

To overcome this, the method utilizes a Zorbax SPE C18 cartridge to significantly reduce interfering substances before quantitation via reversed-phase HPLC. The text provides specific instructions for sample preparation, including pretreatment and washing steps, as well as details on the internal standard used (nitrobenzene) and the final HPLC conditions.

Read online

Agilent Zorbax SPE C18 specifications

Cartridge Capacity100 mg, 1 ml
Part Number5184-3590
Column TypeZorbax SB-C8, 4.6 mmx 150 mm, 3.5 pm
Mobile Phase Composition50% Acetonitrile in 25 mM Ammonium acetate
Detection Wavelength240 nm
Flow Rate1.0 mL/min.

Source: Agilent Zorbax SPE C18 Cartridge Application Brief for Naproxen Analysis

All Agilent Zorbax SPE C18 documents by type

Questions answered in these manuals

What is the purpose of using nitrobenzene in this analysis?

Nitrobenzene is used as an internal standard for High Load and Low Load samples to ensure accurate quantitation.

From: Agilent Zorbax SPE C18 Cartridge Application Brief for Naproxen Analysis

How does the Zorbax SPE C18 cartridge improve naproxen detection?

It significantly reduces levels of interfering substances, such as blood proteins, allowing the analyte to be clearly detected and quantifiable.

From: Agilent Zorbax SPE C18 Cartridge Application Brief for Naproxen Analysis

What are the volumes and solutions required for conditioning, loading, washing, and eluting the cartridge?

The procedure requires adding 1.0 mL of methanol and 1.0 mL of deionized water to condition the cartridge, followed by loading 1.0 mL of sample, washing with 1.0 mL of deionized water and 1.0 mL of 10% methanol, and finally eluting with 1.0 mL of 1% ammonia in methanol.

From: Agilent Zorbax SPE C18 Cartridge Application Brief for Quinine Analysis

What sample preparation steps are needed to create the high load and low load solutions?

To prepare a high load solution, mix 1 mg/mL quinine with 0.5 mL bovine serum and 0.1 mL deionized water; for a low load solution, use 0.05 mg/mL quinine mixed with the same volumes of serum and water.

From: Agilent Zorbax SPE C18 Cartridge Application Brief for Quinine Analysis

What is the purpose of using a Zorbax SPE C18 cartridge for ranitidine analysis?

It significantly reduces levels of interfering substances, such as blood proteins, allowing the analyte to be easily quantitated.

From: Agilent Zorbax SPE C18 Cartridge Application Brief

What internal standard is used in the described method?

Prednisolone is included in the samples as an internal standard.

From: Agilent Zorbax SPE C18 Cartridge Application Brief