Agilent G2399A Solvent Vapor Exit Accessory Installation and Operation Guide
The document details the installation of the Agilent G2399A Solvent Vapor Exit (SVE) Accessory on a Gas Chromatograph (GC). It covers safety precautions, required tools and parts lists, and step-by-step procedures for preparing the GC, installing the valve fitting assembly, and installing the tri-column assembly. The guide also explains how the SVE works to enable large volume injections by separating solvent from analytes.
Read onlineAll Agilent G2399A documents by type
How to use the product for one specific task, with worked examples.
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
Questions answered in these manuals
What is the maximum injection volume achievable with standard pulsed splitless techniques on Agilent 6890 and 5890 Series GCs?
These systems allow approximately two times the normal injection volume, up to 5 uL depending on the solvent.
From: Large-Volume Injection for Gas Chromatography Using COC-SVE Application Note
What is the recommended delay time for the SVE valve when analyzing a diluted drinking water extract with a 25-uL injection?
The recommended delay time is 0.45 minute.
From: Large-Volume Injection for Gas Chromatography Using COC-SVE Application Note
What are the applications for the SVE accessory?
The SVE is suitable for trace analysis of volatiles and semi-volatiles, trace analysis of pesticides, and complete solvent venting to prolong column and detector life.
From: Agilent G2399A Solvent Vapor Exit Accessory Installation and Operation Guide
What tools are required for installation?
Required tools include an ESD protective wriststrap, T-20 Torx screwdriver, diagonal cutters, open-end wrenches (5/16 x 1/4-inch), X-ACTO knife, column cutter, and a flow meter.
From: Agilent G2399A Solvent Vapor Exit Accessory Installation and Operation Guide