Agilent Zorbax SPE C18 Cartridge Application Brief

The document describes a solid phase extraction method for a pharmaceutical compound called Ranitidine. This compound is used for the treatment and prevention of ulcers and for the relief of heartburn. By using a Zorbax SPE C18 cartridge, levels of interfering substances in the bloodstream can be significantly reduced, allowing for easy quantitation of the analyte of interest using reversed-phase HPLC.


Agilent Zorbax SPE C18 Cartridge Application Brief - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
29 KB
Published
03 July, 2012
Updated
29 September, 2023
MD5 checksum
75EBB8006036F3BFB51EC39AF1492DBA

About this manual

This document presents an application brief for the Zorbax SPE C18 cartridge, demonstrating its effectiveness in reducing matrix interference from blood proteins when analyzing ranitidine. The text illustrates how the cartridge allows for the clear detection and quantitation of the analyte via reversed-phase HPLC.

What's inside

  • Application Brief
  • Pharmaceutical
  • Solid Phase Extraction
  • Sample Preparation
  • HPLC Method
  • Recovery
  • Internal Standard

Specifications

Cartridge Capacity100 mg
Cartridge Volume1 mL
Column Dimensions4.6 mm x 150mm
Column Particle Size5um
Flow Rate1.0 mL/min
Column Temperature35°C
Injection Volume20 uL
Detection Wavelength220 nm
Recovery Rate100%

Frequently asked questions

What is the purpose of using a Zorbax SPE C18 cartridge for ranitidine analysis?

It significantly reduces levels of interfering substances, such as blood proteins, allowing the analyte to be easily quantitated.

What internal standard is used in the described method?

Prednisolone is included in the samples as an internal standard.

What is the recommended recovery rate for ranitidine in this method?

The recovery rate for ranitidine is 100%.