Agilent Eclipse Plus C18 Manuals

ManualLib has 16 documents for the Agilent Eclipse Plus C18 (Other Electronics): 9 application notes, 1 product leaflet, 5 datasheets. Every file can be read online or downloaded as a PDF for free.

Application Note 9Product Leaflet 1Datasheet 5

Key facts

Brand
Agilent
Model
Eclipse Plus C18
Documents
16
Languages
JA · EN
Last added
2012-07-06

Start here

Agilent ZORBAX Eclipse Plus C18 Application Note for Simvastatin Analysis

Agilent ZORBAX Eclipse Plus C18 Application Note for Simvastatin Analysis

Application Note 12 pages · 753 KB · JA

This document details the optimization of a USP method for analyzing simvastatin tablets. The original method utilizes an Agilent ZORBAX Eclipse Plus C18 5 µm column (250 mm x 4.6 mm) with a 10-minute analysis time.

The note demonstrates how to modify this method using smaller particle sizes and shorter columns, specifically the Rapid Resolution Eclipse Plus C18 (3.5 µm) and Rapid Resolution HT Eclipse Plus C18 (1.8 µm). These modifications significantly reduce analysis time and solvent usage while maintaining performance requirements. This manual is written in Japanese.

Read online

Agilent Eclipse Plus C18 specifications

Original Column Part Number959990-902
Rapid Resolution Column Dimensions4.6 mm x 100 mm, 3.5 µm
Flow Rate1.5 mL/min
Column Temperature45 °C
Detection Wavelength238 nm

Source: Agilent ZORBAX Eclipse Plus C18 Application Note for Simvastatin Analysis

All Agilent Eclipse Plus C18 documents by type

Questions answered in these manuals

What is the analysis time for the original USP method?

The original USP method requires a 10-minute analysis time.

From: Agilent ZORBAX Eclipse Plus C18 Application Note for Simvastatin Analysis

What is the required flow rate for the Simvastatin analysis method?

The method requires a flow rate of 1.5 mL/min.

From: Agilent ZORBAX Eclipse Plus C18 Application Note for Simvastatin Analysis

What is the maximum pressure limit for the Agilent Poroshell 120 EC-C18 column?

The maximum pressure limit for the Agilent Poroshell 120 EC-C18 column is 356 bar.

From: Agilent Poroshell 120 EC-C18 and ZORBAX Eclipse Plus C18 Method Transfer Technical Overview

What is the flow rate used in the generic gradient analysis described in the document?

The flow rate used in the generic gradient analysis is 2.0 mL/min.

From: Agilent Poroshell 120 EC-C18 and ZORBAX Eclipse Plus C18 Method Transfer Technical Overview

What is the maximum operating pressure for these columns?

The columns support operation up to 1200 bar (17,000 psi).

From: Agilent ZORBAX RRHD Eclipse Plus C18 Threaded Column Datasheet

What is the recommended pH range for use?

These columns can be used over a pH range of 2–9.

From: Agilent ZORBAX RRHD Eclipse Plus C18 Threaded Column Datasheet