Agilent Zorbax SPE C18 Cartridge Application Brief for Naproxen Analysis

This document describes a method for determining naproxen levels in serum using a Zorbax SPE C18 cartridge. The method uses solid phase extraction to significantly reduce the levels of interfering substances. The target analyte is then quantified by reversed-phase HPLC.


Agilent Zorbax SPE C18 Cartridge Application Brief for Naproxen Analysis - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
35 KB
Published
03 July, 2012
Updated
30 September, 2023
MD5 checksum
BB22E1D13C90CFCAEE38781408AFC0FA

About this manual

The document details an analytical procedure for determining levels of naproxen, a pharmaceutical compound used for pain relief, in blood serum. The primary challenge addressed is the interference caused by high protein content in the sample matrix.

To overcome this, the method utilizes a Zorbax SPE C18 cartridge to significantly reduce interfering substances before quantitation via reversed-phase HPLC. The text provides specific instructions for sample preparation, including pretreatment and washing steps, as well as details on the internal standard used (nitrobenzene) and the final HPLC conditions.

What's inside

  • Application Brief
  • Sample Preparation
  • Solid Phase Extraction Method
  • HPLC Method
  • Recovery Results

Specifications

Cartridge Capacity100 mg, 1 ml
Part Number5184-3590
Column TypeZorbax SB-C8, 4.6 mmx 150 mm, 3.5 pm
Mobile Phase Composition50% Acetonitrile in 25 mM Ammonium acetate
Detection Wavelength240 nm
Flow Rate1.0 mL/min.

Frequently asked questions

What is the purpose of using nitrobenzene in this analysis?

Nitrobenzene is used as an internal standard for High Load and Low Load samples to ensure accurate quantitation.

How does the Zorbax SPE C18 cartridge improve naproxen detection?

It significantly reduces levels of interfering substances, such as blood proteins, allowing the analyte to be clearly detected and quantifiable.

What is the recovery rate for high load samples?

The recovery for a high load (10 pg/ml) was 99.25%.