Agilent 1100 Series GPC-SEC System Characterization of Light-Emitting Polymers Application Note
This document describes a method for characterizing light-emitting polymers, specifically phenyl-substituted poly(p-phenylene vinylene) (PPV), to monitor quality parameters such as molecular weight and purity. The active layer of polymer-LEDs can be prepared by simple coating methods like spin coating.
Read onlineAgilent 1100 Series GPC-SEC system specifications
| Column | 3 x PLgel mixed B, 7.5 x 300 mm, 10 um |
|---|---|
| Mobile phase | Tetrahydrofuran |
| Flow rate | 1.0 mL/min |
| Column compartment temperature | 20°C |
| Injection volume | 10 ul |
| Detector | Refractive index detector, alternatively VWD, 254 nm |
Source: Agilent 1100 Series GPC-SEC System Process Control of Polystyrenes Application Note
All Agilent 1100 Series GPC-SEC system documents by type
How to use the product for one specific task, with worked examples.
Questions answered in these manuals
What is the concentration of the samples used for the analysis?
The samples were dissolved at a concentration of 0.1 %.
From: Agilent 1100 Series GPC-SEC System Characterization of Light-Emitting Polymers Application Note
What system is required to characterize small differences in polymers reliably?
A GPC-SEC instrument with excellent precision, such as the Agilent 1100 Series GPC-SEC system.
From: Agilent 1100 Series GPC-SEC System Process Control of Polystyrenes Application Note
What standards were used for narrow standard calibration?
Polystyrene EasyCal Vial standards (Agilent p/n 79911-60500 and 79911-60501) were used.
From: Agilent 1100 Series GPC-SEC System Process Control of Polystyrenes Application Note