Agilent J&W DB-5ms Ultra Inert GC Column PBDE Analysis Application Note

This application demonstrates the use of electron impact GC–MS for trace-level PBDE analysis. The method uses Agilent J&W DB-5ms Ultra Inert GC column. This column has been thoroughly tested for column inertness and provides the analyst with high confidence in the accuracy of the results.


Agilent J&W DB-5ms Ultra Inert GC Column PBDE Analysis Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
53 KB
Published
11 June, 2012
Updated
30 September, 2023
Model
DB-5MS
MD5 checksum
177090A4DCCD5E00DF70018C17CCDB4D

About this manual

This document describes a method for analyzing polybrominated diphenyl ethers (PBDEs) using electron impact gas chromatography-mass spectrometry (GC-MS). The application highlights the use of the Agilent J&W DB-5ms Ultra Inert capillary GC column to achieve successful separation of challenging analytes, particularly BDE-209, by minimizing column activity and residence time at high temperatures.

The study details the chromatographic conditions, including the use of a 15-meter column length to limit thermal exposure and a specific temperature program. Results indicate excellent linearity with R² values of 0.997 or greater and a signal-to-noise ratio greater than three to one for BDE-209 at a loading of 0.025 ng.

What's inside

  • Introduction
  • PBDE Analyses
  • Chromatographic Conditions
  • Results
  • System Configuration

Specifications

Column Length15 m
Column Dimensions15 m x 0.25 mm, 0.25 µm
Column Part Number122-5512UI
Carrier GasHelium
Inlet Temperature325 °C
Oven Temperature Range150 to 325 °C
Oven Ramp Rate17 °C/min
Oven Hold Time5 min
MSD Source Temperature300 °C
Quadrupole Temperature150 °C
Transfer Line Temperature300 °C
Scan Range200-1000 amu
Linearity R²0.997 or greater

Frequently asked questions

What column length was used for the PBDE analysis?

A 15-meter long column was used to limit residence time for BDE-209.

What is the signal-to-noise ratio for BDE-209 at 0.025 ng loading?

The signal-to-noise ratio was greater than three to one.

What is the part number for the Agilent J&W DB-5ms Ultra Inert column used?

The part number is 122-5512UI.

Download PDF (53 KB)

More Agilent Manuals