Agilent J&W DB-5ms Ultra Inert Capillary GC Column Application Brief for PAH Analysis

This document describes the results of trace-level PAH analyses using Agilent J&W DB-5ms Ultra Inert 30 m × 0.25 mm, 0.25 µm capillary GC column on a 16-component PAH standard mixture.


Agilent J&W DB-5ms Ultra Inert Capillary GC Column Application Brief for PAH Analysis - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
131 KB
Published
11 June, 2012
Updated
01 October, 2023
Model
DB-5MS
MD5 checksum
B2DB8C5E63928E223F1FCA8341E8D436

About this manual

This document describes a method for analyzing trace- and ultra trace-level polycyclic aromatic hydrocarbons (PAHs). The application utilizes electron impact single quadrupole scanning mass spectrometry with an Agilent J&W DB-5ms Ultra Inert capillary GC column to achieve excellent sensitivity and resolution.

What's inside

  • Introduction
  • Agilent Testing Procedure
  • PAH Analysis Results
  • Chromatographic Conditions
  • Column Specifications

Specifications

Column Length30 m
Column Inner Diameter0.25 mm
Film Thickness0.25 µm
Part Number122-5532UI

Frequently asked questions

How long is the analysis time for the PAH mixture?

Good resolution was obtained in a 15-min analysis for each of the PAHs.

What is the minimum concentration range demonstrated for PAH analysis?

The application demonstrates trace-level analyses over a concentration range of 0.05 pg/mL to 5 pg/mL.

How long does the analysis take for each PAH component?

Good resolution was obtained in a 15-min analysis for each of the PAHs.

What is the minimum column loading that still yields excellent linearity?

Excellent linearity was achieved down to a 0.05 ng column loading of each component.