Agilent J&W DB-5ms Ultra Inert Capillary GC Column PBDE Analysis Application Note

This application demonstrates the trace-level analysis of polybrominated diphenyl ethers (PBDE) using electron impact single quadrupole scanning mass spectrometry. Agilent Technologies Inc. has implemented new testing procedures to evaluate the column inertness performance more effectively. The study focuses on the importance of PBDE analysis in understanding food supply and environmental quality worldwide.


Agilent J&W DB-5ms Ultra Inert Capillary GC Column PBDE Analysis Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
6
File format
PDF
File size
127 KB
Published
02 July, 2012
Updated
28 September, 2023
MD5 checksum
3B023E2119342F0287718A6A04F3AB21

About this manual

Polybrominated diphenyl ethers (PBDEs) are persistent environmental chemicals used as flame retardants. This document details a method for analyzing trace and ultra-trace levels of PBDE congeners, including the challenging BDE-209, using gas chromatography-mass spectrometry (GC/MS).

The application highlights the use of the Agilent J&W DB-5ms Ultra Inert 15 m x 0.25 mm x 0.25 µm capillary GC column (part number 122-5512UI). The method utilizes an Agilent 6890N GC/5973B MSD with a 7683B autosampler and specific chromatographic conditions to achieve excellent sensitivity and linearity for various BDE congeners.

What's inside

  • Introduction
  • Sample Preparation
  • Chromatographic Conditions
  • Flow Path Supplies
  • Baseline Inertness Profile
  • PBDE Analyses Results

Specifications

Column Length15 m
Column Inner Diameter0.25 mm
Column Film Thickness0.25 µm
Column Part Number122-5512UI
Carrier Gas Flow Rate72 cm/s
Inlet Temperature325 °C
Oven Temperature Range150 to 325 °C
MSD Source Temperature300 °C

Frequently asked questions

Which Agilent column is recommended for challenging PBDE analysis?

The document recommends the Agilent J&W DB-5ms Ultra Inert capillary GC column.

What are the concentration ranges used for BDE calibration?

The study used a range of 0.5 to 250 ng/mL for most BDEs and 2.5 to 1,000 ng/mL for BDE-209.