Agilent Mx3000P/Mx3005P QPCR System Instrument Qualification Beta-Actin Install Plate Instructions

Agilent Mx3000P/Mx3005P QPCR System Instrument Qualification Beta-Actin Install Plate Instructions

User Manual 2 pages · 192 KB · EN

This document provides instructions for the storage, handling, and preparation of the Beta-Actin Install Plate used with the Agilent Mx3000P or Mx3005P QPCR System. It details procedures such as thawing, mixing by inversion, centrifugation, and bubble removal to ensure accurate qualification testing.

Read online

Agilent MX3000P specifications

Reaction Timeunder 50 minutes
CompatibilityMx3000P and Mx3005P
Amplification Efficiency96.4%
Standard Curve R20.999
RT Synthesis Temperature50°C
Packaging Qty2 x 2 ml

Source: Agilent Brilliant III Ultra-Fast QPCR/QRT-PCR Master Mixes for Mx3005P/Mx3000P handbook

All Agilent MX3000P documents by type

User Manual 3

Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.

Datasheet 1

Ratings, dimensions and ordering codes, for selection and design work.

Questions answered in these manuals

What is the maximum reaction time for Brilliant III Ultra-Fast reagents?

The reagents allow the completion of real-time experiments in less than 50 minutes on the Mx platform.

From: Agilent Brilliant III Ultra-Fast QPCR/QRT-PCR Master Mixes for Mx3005P/Mx3000P handbook

Which Agilent instruments are validated for these reagents?

The reagents have been validated on the Agilent Mx3000P and Mx3005P instruments.

From: Agilent Brilliant III Ultra-Fast QPCR/QRT-PCR Master Mixes for Mx3005P/Mx3000P handbook

What four foodborne pathogens can be detected using these assays?

The assays are designed to detect Salmonella enterica, Escherichia coli O157:H7, Listeria monocytogenes, and Campylobacter jejuni.

From: 5'-Nuclease PCR Assays for Foodborne Pathogen Detection using Agilent-Stratagene Mx3000P Q-PCR System

How long does it take to obtain results after sample enrichment?

Results can be obtained with these assays in 2.5 h following enrichment.

From: 5'-Nuclease PCR Assays for Foodborne Pathogen Detection using Agilent-Stratagene Mx3000P Q-PCR System

What is the linear detection range of the multiplex assay for mycotoxigenic fungi?

The linear detection range of the multiplex assay is between 10 pg to 10 ng of DNA.

From: Agilent Mx3000P QPCR System Application Note for Mycotoxigenic Fungi Detection

What are the optimum parameters for primer and probe concentration in the PCR reaction?

The optimum parameters are 10 pmol of each primer and 5 pmol of probe per reaction.

From: Agilent Mx3000P QPCR System Application Note for Mycotoxigenic Fungi Detection