Agilent 255 NCD Manuals

ManualLib has 4 documents for the Agilent 255 NCD: 3 product leaflets, 1 service manual. Every file can be read online or downloaded as a PDF for free.

Product Leaflet 3Service Manual 1

Key facts

Brand
Agilent
Model
255 NCD
Documents
4
Languages
EN
Last added
2012-07-05

Start here

Agilent 255 Nitrogen Chemiluminescence Detector (255 NCD) Product Information

Agilent 255 Nitrogen Chemiluminescence Detector (255 NCD) Product Information

Product Leaflet 2 pages · 36 KB · EN

The Agilent 255 Nitrogen Chemiluminescence Detector (255 NCD) is designed to easily detect organic compounds containing nitrogen. It achieves this by converting the compounds into nitric oxide using a stainless steel burner that operates in a hydrogen and oxygen plasma at temperatures greater than 1800 °C.

The detector utilizes a red optical filter to transmit light from the chemiluminescence reaction while suppressing signals from other chemical species. This selectivity ensures that major constituents of sample matrices, such as carbon dioxide, water, nitrogen gas, and noble gases, do not interfere significantly with the determination of trace levels of nitrogen-containing analytes.

Read online

All Agilent 255 NCD documents by type

Service Manual 1

Disassembly, diagnostics and part numbers, for repair.

Questions answered in these manuals

What compounds can the Agilent 255 NCD detect?

The detector can detect organic compounds containing nitrogen after they are converted to nitric oxide, as well as inorganic compounds like ammonia and hydrazine.

From: Agilent 255 Nitrogen Chemiluminescence Detector (255 NCD) Product Information

What is the operating temperature of the burner?

The stainless steel burner operates at temperatures greater than 1800 °C to convert nitrogen compounds to nitric oxide.

From: Agilent 255 Nitrogen Chemiluminescence Detector (255 NCD) Product Information

What is the run time for the Agilent 255 NCD when using 0.53 mm id column conditions?

The run time for the Agilent 255 NCD is 10 minutes.

From: Agilent 255 NCD Nitrosamines Analysis Application Manual

How often must the CTR traps be replaced when used with the Agilent 255 NCD?

The CTR traps require replacement after approximately 50 analyses.

From: Agilent 255 NCD Nitrosamines Analysis Application Manual

What is the principle of operation for the 355 SCD?

The detector utilizes the combustion of sulfur compounds to form sulfur monoxide (SO) and the subsequent chemiluminescence reaction of SO with ozone (O3).

From: Agilent 355 SCD and 255 NCD Dual Plasma Chemiluminescence Detectors Product Leaflet

What are the major applications for the 255 NCD?

The detector is used for chemicals, environmental samples, food and beverages, fuels, gases, pesticides and herbicides, petrochemicals, polymers, and nitrosamines.

From: Agilent 355 SCD and 255 NCD Dual Plasma Chemiluminescence Detectors Product Leaflet