Agilent 7890A GC/5975C MSD VOC Analysis Application Note

This application note describes a method for the analysis of volatile organic compounds (VOCs) in water using static headspace-GC/MS, including method optimization, sample preparation, and analytical method.


Agilent 7890A GC/5975C MSD VOC Analysis Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
16
File format
PDF
File size
457 KB
Published
02 July, 2012
Updated
30 September, 2023
MD5 checksum
55B1818AEC86C253670605A2C961A8C2

About this manual

This document describes a static headspace (SHS) method optimized for the determination of volatile organic compounds (VOCs) in water. The method utilizes an Agilent 7890A GC coupled with a 5975C MSD equipped with triple axis detection to achieve purge and trap sensitivities while maintaining the robustness of static headspace.

What's inside

  • Introduction
  • Sample Preparation
  • Instrumental Conditions
  • Experimental Conditions
  • Column Selection
  • Method Development
  • Target Analytes

Specifications

Sample Volume10-mL water samples
Vial Type20-mL headspace vial
Internal Standard Concentration800 ng/L (800 ppt)
Target Analytes60 target analytes
Calibration Levels45 to 1,250 ng/L
Column TypeDB-624, 20 m x 0.18 mm
Carrier GasHe, constant pressure (95 kPa)
Run Time25-min run

Frequently asked questions

What system was used for the analysis?

The analyses were performed on an Agilent 7890A GC/5975C MSD system.

What reference water was used for calibration?

Bottled drinking water (Evian) was used for reference water blanks.