Agilent GF-250 Manuals

ManualLib has 8 documents for the Agilent GF-250: 7 user manuals, 1 datasheet. Every file can be read online or downloaded as a PDF for free.

User Manual 7Datasheet 1

Key facts

Brand
Agilent
Model
GF-250
Industry
Industrial
Documents
8
Languages
EN
Last added
2012-07-04

Start here

Agilent Zorbax PrepHT GF-250 and GF-450 Datasheet

Agilent Zorbax PrepHT GF-250 and GF-450 Datasheet

Datasheet 2 pages · 106 KB · EN

The document describes the Agilent Zorbax PrepHT GF-250 and GF-450 as surface-stabilized, hydrophilic gel filtration columns designed for the size separation of biological macromolecules. These columns offer superior efficiency and pH stability when used with aqueous buffer solutions in the pH range of 3.0 to 8.5.

Read online

Agilent GF-250 specifications

Optimal Separation Range (GF-450)900,000 to 25,000 Daltons
Optimal Separation Range (GF-250)250,000 to 10,000 D
Recommended pH LimitspH 3 to 8.5
Maximum Operating Temperature40°C
Protein Recovery85-100%

Source: Agilent Zorbax PrepHT GF-250 and GF-450 Datasheet

All Agilent GF-250 documents by type

Datasheet 1

Ratings, dimensions and ordering codes, for selection and design work.

Questions answered in these manuals

What is the recommended pH range for these columns?

The recommended pH limits for these columns are pH 3 to 8.5.

From: Agilent Zorbax PrepHT GF-250 and GF-450 Datasheet

Can the columns be used with organic solvents?

Yes, the GF-250 and GF-450 columns are compatible with most organic solvents.

From: Agilent Zorbax PrepHT GF-250 and GF-450 Datasheet

What is the recommended flow rate for GF-250 columns?

Recommended flow rates are from 0.5-3.0 mL/min, typically 1.0 mL/min.

From: Agilent Zorbax Bio Series GF-250 Datasheet

Can GF-250 columns be used at high pH?

Yes, the columns may be used for short periods at pH 8.5 due to zirconia stabilization.

From: Agilent Zorbax Bio Series GF-250 Datasheet

What is the purpose of using denaturing mobile phases in SEC?

Denaturing mobile phases containing guanidine hydrochloride or SDS are used to insure un-aggregated, monomers for proper size determination.

From: Agilent ZORBAX GF-250 Application Note: Comparison of Denaturing and non-Denaturing Mobile Phases

How does denaturation affect the separation results?

Denaturation shifts molecules toward larger apparent size and smaller retention volume.

From: Agilent ZORBAX GF-250 Application Note: Comparison of Denaturing and non-Denaturing Mobile Phases