This document introduces ZORBAX GF-250 column in the separation of Bio-Rad molecular weight markers
This document describes the application of size-exclusion chromatography to separate molecules based on their hydrodynamic volume. It details a specific method for calibrating and testing ZORBAX GF-250 or GF-450 columns using Bio-Rad gel-filtration markers.
The procedure utilizes a 9.4 x 250 mm ZORBAX GF-250 column with 200mM Sodium Phosphate mobile phase at pH 7.0. The method highlights the high efficiency of the porous-silica microspheres, allowing for excellent resolution of five protein standards ranging from 1.35 Kdal to 670 Kdal.
The document specifies a column size of 9.4 x 250 mm.
Use 200mM Sodium Phosphate at pH 7.0 as the mobile phase.
The procedure requires a 5pL injection with a flow rate of 2 mL/min.