Agilent ZORBAX GF-250 Column Calibration Using Bio-Rad Gel-Filtration Standards Application Note

This document describes how to calibrate a ZORBAX GF-250 column using Bio-Rad gel filtration standards.


Agilent ZORBAX GF-250 Column Calibration Using Bio-Rad Gel-Filtration Standards Application Note - cover page
Brand
Agilent
Document type
Product Leaflet
Language
English
Pages
2
File format
PDF
File size
145 KB
Published
03 July, 2012
Updated
30 September, 2023
MD5 checksum
260E70E25CC4BB4737462AFC30219FFC

About this manual

This document describes a method to calibrate and test the proper operation of the Agilent ZORBAX GF-250 size-exclusion column. It utilizes molecular-weight markers from Bio-Rad gel-filtration standards, which allow calibration in the appropriate molecular-weight range for this specific column type.

The application highlights that these protein standards are separated by size with good linearity. The text notes that while the ZORBAX GF-250 achieves better separation of smaller molecules down to 1 Kdal, the GF-450 is designed for larger molecules up to 1 mdal.

What's inside

  • Application
  • Size-exclusion columns
  • Molecular-weight markers
  • Bio-Rad gel-filtration standards
  • ZORBAX GF-250
  • Retention Volume
  • Mobile Phase
  • UV Detection

Specifications

Column Part Number884973-901
Column Dimensions250 mm length, 9.4.x ID
Mobile Phase Concentration200mM Sodium Phosphate
Flow Rate2 mL/min
Detection Wavelength254 nm
Temperature ConditionAmbient

Frequently asked questions

What standards are used to calibrate the ZORBAX GF-250 column?

Bio-Rad gel-filtration standards are used for calibration.

What is the upper limit of separation for the ZORBAX GF-450?

The GF-450 achieves better separation of large molecules up to 1 mdal.