Agilent 500-MS IT Mass Spectrometer MS Workstation V6 Software Operation Manual
This document serves as an operation manual for the Varian (now Agilent) MS Workstation Version 6.9 software, which controls the 500-MS ion trap mass spectrometer, associated chromatography modules, and the LC Combi PAL AutoSampler. The software manages hardware control, data acquisition, display, reduction, and analysis.
Read onlineAgilent 500-MS specifications
| Mass range | 50 to 2000u |
|---|---|
| Resolution | <0.5u FWHM over the entire range |
| Mass axis stability | + 0.1u over 24 hours at constant temperature |
| Max ion time | 500,000 psec |
| Drying gas temperature (ESI) | 400°C max |
| LC flow range (ESI) | 1 to 1000 pL/min |
| Needle voltage (ESI) | 6 kV max |
| LC flow range (APCI) | 100 pL/min to 2 mL/min |
| Auxiliary gas temperature (APCI) | 550°C max |
| Power requirements | 100-120, 200-240 Vac, +10 %, 50/60 Hz, 3200 VA (steady state) |
| Operating Temperature | 16°C to 30°C |
| Weight (MS) | 135 lb/ 62 kg |
Source: Varian 500-MS LC Ion Trap Mass Spectrometer Specifications
All Agilent 500-MS documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
How to use the product for one specific task, with worked examples.
- Agilent Identification Characterization of a Novel Delivery Molecule for a Clandestine PDE-5 Inhibitor Drug in a Dietary Supplement Application Note 5 p
- Varian 500-MS Ion Trap Mass Spectrometer Application Note 02101 2 p
- Agilent Analysis of Vulcanized Rubber Formulations on the 500-MS Ion Trap Mass Spectrometer Confirmation of Expected Structures Using TurboDDS with ACD/MS Manager Software Application Note
- Varian 500-MS Phentolamine Adulteration in Dietary Supplements Application Note 4 p
- Varian 500-MS Ion Trap Mass Spectrometer Application Note 02126 3 p
- Agilent 500-MS Ion Trap LC/MS Application Note for Ethomeen T/20H Characterization 2 p
- Agilent 500-MS Ion Trap LC/MS Analysis of Isocyanate Derivatives Application Note 2 p
- Agilent 500-MS LC Ion Trap Mass Spectrometer Mass ID Application Note 4 p
Mounting, wiring and first power-up. Read it before the hardware goes in.
Ratings, dimensions and ordering codes, for selection and design work.
Questions answered in these manuals
What is the mass range of the Varian 500-MS?
The mass range is from 50 to 2000u.
From: Varian 500-MS LC Ion Trap Mass Spectrometer Specifications
Can the drying gas temperature be programmed?
Yes, SelecTemp allows temperature programming of the drying gas throughout the analysis.
From: Varian 500-MS LC Ion Trap Mass Spectrometer Specifications
What is the mass range of the 500-MS?
The mass range is 50 to 2000u.
From: Agilent 500-MS LC Ion Trap Mass Spectrometer Specifications
What is the resolution of the analyzer?
The resolution is less than 0.5u FWHM over the entire range.
From: Agilent 500-MS LC Ion Trap Mass Spectrometer Specifications
What is the main difference between an ion trap mass spectrometer and a triple quadrupole mass filter?
Ion traps store ions simultaneously within trapping electrodes and eject them sequentially, allowing MS/MS in the same space. Triple quadrupoles use separate filters and collision cells distributed in space.
What are the strengths of an ion trap mass spectrometer?
They offer fast, sensitive full scan spectra for MS and MS/MS, excellent high mass sensitivity, structural elucidation capabilities via MSn, and a lower price point.