Agilent 6890Plus/5973N Nitrofen Determination Application Note
This document describes an analytical method for determining the herbicide Nitrofen in turkey meat using Gas Chromatography/Mass Spectrometry (GC/MS). The application focuses on the advantages of Negative Chemical Ionization (NCI) mode over Electron Impact (EI) mode for detecting analytes with high electron affinity in complex food matrices.
The text details the experimental setup, including the use of Agilent 6890Plus and 5973N MSD instruments with EPC modules and Retention Time Locking (RTL). It outlines the sample preparation procedure according to DFGS 19 and presents results comparing EI and NCI detection modes, highlighting the superior sensitivity and selectivity of NCI for Nitrofen detection. This manual is written in German.
Agilent 6890 Plus specifications
| Gas Chromatograph | Agilent 6890Plus |
|---|---|
| Mass Spectrometer | Agilent 5973N, MSD, CL-Option |
| Injection Volume | 1uL, Pulsed Pressure Splitless |
| NCI Moderator Gas | Methan |
| Ion Source Temperature | 150°C |
Source: Agilent 6890Plus/5973N Nitrofen Determination Application Note
All Agilent 6890 Plus documents by type
How to use the product for one specific task, with worked examples.
Questions answered in these manuals
Which detection mode is preferred for Nitrofen?
Negative Chemical Ionization (NCI) mode is preferred because Nitrofen shows high electron affinity.
From: Agilent 6890Plus/5973N Nitrofen Determination Application Note
What is the detection limit for Nitrofen in a turkey meat matrix under the specified NCI SIM conditions?
The detection limit for Nitrofen in the turkey meat matrix is 200 fg.
From: Agilent 6890Plus/5973N Nitrofen Determination Application Note
What is the typical range for retention time reproducibility on this GC?
A given peak typically falls within a very tight range of 0.002 to 0.003 min.
From: Agilent 6890 Plus GC Retention Time Reproducibility Technical Note
What is the average standard deviation for all tested GCs?
The average standard deviation for all GCs tested was 0.001 min.
From: Agilent 6890 Plus GC Retention Time Reproducibility Technical Note
What is the maximum oven temperature for the Agilent 6890N, 6890 Plus, and 4890D models?
The maximum oven temperature for these models is 450 °C.
What is the maximum number of channels for the Agilent 6890N, 6890 Plus, and 4890D models?
The maximum number of channels for these models is 2.