Varian 500-MS Ion Trap Mass Spectrometer Application Note 02126

This document describes a method for analyzing vulcanized rubber formulations using the Varian 500-MS ion trap mass spectrometer and confirming expected structures using Varian TurboDDS™ and ACD/MS Manager software.


Varian 500-MS Ion Trap Mass Spectrometer Application Note 02126 - cover page
Brand
Agilent
Category
Other Electronics
Document type
Application Note
Language
English
Pages
3
File format
PDF
File size
255 KB
Published
12 June, 2012
Updated
30 September, 2023
Model
500-MS
MD5 checksum
8F5E07401F9A301DA994063D160FB939

About this manual

This document details a method for analyzing vulcanized rubber extracts to identify polymeric components and specific chemical additives, such as Antioxidant 2246. The process involves liquid chromatography/mass spectrometry (LC/MS) to separate compounds and mass spectrometry/mass spectrometry (MS/MS) for structural elucidation.

The analysis utilizes the Varian 500-MS Ion Trap Mass Spectrometer equipped with TurboDDS software to automatically trigger MS/MS data collection for prominent peaks. The resulting data is then processed using ACD/MS Manager software to match predicted fragmentation patterns against experimental spectra, confirming the identity of suspected compounds.

What's inside

  • Introduction
  • Materials and Reagents
  • Sample Preparation
  • HPLC Conditions
  • MS Conditions
  • TurboDDS Scan Functions
  • Instrumentation
  • Results and Discussion
  • Structural Analysis
  • Conclusion

Specifications

ColumnPursuit XRs C18, 3 pm, 150 x 3.0 mm
Mobile PhaseAcetonitrile
LC Flow Rate200 pL/min
LC Run Time30 min
Injection Volume20 pL
Nebulizing Gas35 psi
Drying Gas10 psi at 350 °C
Needle Voltage5000 V
Capillary Voltage80V
Shield Voltage600 V
RF Loading100%
Detector Voltage1900 V
Survey Scan Rangem/z 100-700
MS/MS Trigger Threshold8600 counts
Minimum Intensity Ratio1%
Isotopic Equivalence3 m/z
Full Scan Rangem/z 100-1000

Frequently asked questions

What is the m/z value of the suspected chemical component Antioxidant 2246?

The suspected chemical component, Antioxidant 2246, has an m/z value of 341 for the protonated molecule.

What is the specific purpose of this application note?

This document describes how to analyze vulcanized rubber formulations and confirm expected structures using the Varian 500-MS Ion Trap Mass Spectrometer with TurboDDS software.

What are the HPLC flow rate and run time settings for this method?

The liquid chromatography program uses an isocratic flow at 200 µL/min for a total run time of 30 minutes.

What is the mass range scanned during the survey scan?

The survey scan function collects data over a mass-to-charge ratio (m/z) range of 100 to 700.