This application note describes large-volume injection (LVI) with a cool on-column (COC) inlet and solvent vapor exit (SVE) kit and is illustrated with examples of straight-chain hydrocarbons, polycyclic aromatic hydrocarbons (PAHs), and semivolatiles.
The document explains that Large-Volume Injection (LVI) is a valuable tool for trace analysis by gas chromatography, enabling injection volumes up to 1 mL. It focuses on the Cool On-Column (COC) inlet combined with a Solvent Vapor Exit (SVE) kit to selectively evaporate bulk solvent before sample transfer.
The text details the hardware configuration involving a precolumn assembly and vent line, as well as the software integration within Agilent ChemStation for automated method development. It provides experimental results demonstrating the technique's effectiveness for analyzing straight-chain hydrocarbons, polycyclic aromatic hydrocarbons (PAHs), and semivolatiles in drinking water extracts.
These systems allow approximately two times the normal injection volume, up to 5 uL depending on the solvent.
The recommended delay time is 0.45 minute.
ChemStation software version 4.04.02 or higher is required.