Key facts
- Brand
- Agilent
- Model
- 7500CE
- Industry
- Industrial
- Documents
- 12
- Languages
- EN
- Last added
- 2012-07-05
Start here
Agilent 7500CE specifications
| Detection limit (in methanol) | < 1 pg per compound |
|---|---|
| Detection limit (in biogas) | 0.03 - 0.07 ppbv |
| Total run time | less than 10 minutes |
Source: Agilent 6890/7500ce GC-ICP-MS Siloxanes Analysis Application Note
All Agilent 7500CE documents by type
Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.
How to use the product for one specific task, with worked examples.
- Agilent 7500ce ICP-MS Application Note: Determination of 18 Elements in Blood Serum 8 p
- Agilent 7500ce ORS ICP-MS Carbonate Associated Sulfur Analysis Application Note 4 p
- Achieving Optimum Throughput in ICP-MS Analysis with the Agilent 7500ce ICP-MS 6 p
- Agilent 7500ce ORS ICP-MS Direct Elemental Analysis of Gasoline Application Note 8 p
- Agilent 7500ce Multi-Element Determination of Heavy Metals in Dietary Supplements Using Collision/Reaction Cell ICP-MS A 4 p
- Agilent 7500ce ICP-MS Unmatched Spectral Interference Removal Using ORS with Helium Collision Mode Application Note 8 p
- Agilent 7500ce ICP-MS Chromium Speciation Application Note 8 p
- Agilent 7500ce ICP-MS Application Note: The ORS Advantage for High Matrix Analysis 12 p
Questions answered in these manuals
What are the detection limits for siloxanes in biogas using this method?
The resulting detection limits ranged from 0.03 to 0.07 ppbv.
From: Agilent 6890/7500ce GC-ICP-MS Siloxanes Analysis Application Note
How does the detection limit compare to GC/MS?
The GC-ICP-MS method is more than 500 times more sensitive than the lowest reported GC/MS detection limit.
From: Agilent 6890/7500ce GC-ICP-MS Siloxanes Analysis Application Note
What types of samples is the Agilent 750ce ICP-MS designed to analyze?
The instrument is designed for analyzing environmental, clinical, and other high matrix samples such as wastewater, seawater, soils, foods, and biological samples.
What gases are used by the Octopole Reaction System for interference removal?
The system uses simple cell gases consisting of hydrogen and helium to remove interfering species.
What sample preparation is required for this method?
The only sample preparation necessary was dilution using an alka-line diluent containing ammonia, EDTA Triton X-100, and butan-1-ol.
From: Agilent 7500ce ICP-MS Application Note: Determination of 18 Elements in Blood Serum
Which gas modes were used for the analysis?
The Agilent 7500ce was used in three different gas modes: hydrogen, helium, and standard or no-gas mode.
From: Agilent 7500ce ICP-MS Application Note: Determination of 18 Elements in Blood Serum