Agilent 7500ce Multi-Element Determination of Heavy Metals in Dietary Supplements Using Collision/Reaction Cell ICP-MS A

The study determined 11 metals in dietary supplements consumed in the UK using ICP-MS. The method was validated by analyzing two certified reference materials and showed accurate and reliable results.


Agilent 7500ce Multi-Element Determination of Heavy Metals in Dietary Supplements Using Collision/Reaction Cell ICP-MS A - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
79 KB
Published
12 June, 2012
Updated
29 September, 2023
Model
7500CE
MD5 checksum
81B1C6D45A4C14D1BF1ECF0D76989F6D

About this manual

The document details a method for analyzing 11 specific metals, including arsenic, chromium, mercury, iron, copper, nickel, zinc, selenium, lead, cadmium, and thallium, in dietary supplements consumed in the UK. The analysis was performed using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) equipped with an Octopole Reaction System (ORS).

What's inside

  • Abstract
  • Introduction
  • Experimental
  • Sample Preparation
  • QC Materials
  • Instrumentation
  • Measurement
  • Results and Discussion

Specifications

Instrument ModelAgilent 7500ce CRC-ICP-MS
Cell SystemOctopole Reaction System (ORS)
Number of Elements Analyzed11 metals
Microwave Program Temperature180 °C

Frequently asked questions

Which elements were analyzed in this study?

The study analyzed arsenic, chromium, iron, copper, nickel, zinc, selenium, lead, cadmium, and thallium.

What collision/reaction cell modes were used?

The instrument was operated in no-gas, helium, and hydrogen modes.

How many certified reference materials were used for assessment?

Two certified reference materials, LGC7160 and SRM1577b, were analyzed.

What was the recovery range for the certified values?

Recoveries were 100 +/- 15% of the certified value for all elements.

Download PDF (79 KB)