Agilent 7500ce ORS ICP-MS Direct Elemental Analysis of Gasoline Application Note

The Agilent 7500ce ICP-MS is shown to be an excellent analytical tool for the elemental analysis of gasoline. It offers the best detection limits of any multi-element tech- nique and a full elemental analysis suite that takes less than 5 minutes to acquire. The Octopole Reaction System (ORS), integral to the 7500ce, ensures that a single method and set of conditions can be used to directly determine all elements of interest in the sample. Comparison data for selected elements in both gas mode and no-gas mode highlight the effectiveness of the ORS for interference removal. Results showing the excellent long-term stability of the 7500ce ORS are also given.


Agilent 7500ce ORS ICP-MS Direct Elemental Analysis of Gasoline Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
583 KB
Published
02 July, 2012
Updated
30 September, 2023
Models
7500CE, 7500c
MD5 checksum
96F8A9275388D79091CC962BA8EEAE96

About this manual

The document describes the use of the Agilent 7500ce ICP-MS equipped with the Octopole Reaction System (ORS) for the direct analysis of gasoline. It addresses challenges such as high vapor pressure and carbon deposition by utilizing a cooled spray chamber, a narrow bore torch injector, and oxygen addition to the plasma.

The note details the operating conditions, including forward power, gas flows, and the use of platinum interface cones. It explains the application of different ORS modes (Helium, Hydrogen, and No-gas) to remove specific polyatomic interferences and presents calibration data and detection limits for elements like Magnesium, Chromium, Cadmium, and Barium in gasoline samples.

What's inside

  • Abstract
  • Introduction
  • Trace Impurities in Gasoline
  • Sample Preparation and Instrumental Conditions
  • Data Acquisition
  • On-Mass Measurements
  • Results and Discussion

Specifications

Forward power1200 W
Plasma gas flow15.0 L min
Auxiliary gas flow1.0 L min
Nebulizer gas flow0.70 L min
Make up gas flow0.13 L min
Sampling depth9.0 mm

Frequently asked questions

What gas modes are used for data acquisition?

Data is acquired operating the ORS in helium mode, hydrogen mode, and no-gas mode.

How is carbon interference on Magnesium removed?

A slightly higher hydrogen flow rate is used for the removal of C interference on Mg.

What is the purpose of adding oxygen to the plasma?

Oxygen is introduced to eliminate the effects of carbon deposition on the interface and ion lenses.