Agilent 710-ES Manuals

ManualLib has 3 documents for the Agilent 710-ES: 2 user manuals. Every file can be read online or downloaded as a PDF for free.

User Manual 2

Key facts

Brand
Agilent
Model
710-ES
Industry
Industrial
Documents
3
Languages
EN
Last added
2012-06-12

Start here

Agilent 710-ES Determination of Mercury in Certified Reference Sludge Application Note

Agilent 710-ES Determination of Mercury in Certified Reference Sludge Application Note

Application Note 4 pages · 39 KB · EN

The document details the determination of mercury levels in solid wastes, specifically sewage sludges, which often contain high concentrations due to environmental contamination. It highlights the utility of vapor generation techniques coupled with Inductively Coupled Plasma-Optical Emission Spectrometry (ICP-OES) for measuring mercury at low concentrations typically found in environmental analysis.

The note presents a specific procedure using an Agilent 710-ES equipped with a megapixel CCD detector and an Agilent VGA-77P Vapor Generation Accessory. It outlines the operating conditions for both the ICP instrument and the VGA, including gas flows, power settings, and pump rates. The document also covers sample preparation methods involving acid digestion of NBS 2781 (Domestic Sludge) reference material and reports detection limits and analysis results.

Read online

Agilent 710-ES specifications

Detector Typemegapixel CCD detector
Power1.4 kW
Plasma gas flow15 L/min
Auxiliary gas flow1.5 L/min
Spray chamber typeGlass cyclonic
Torch typeStandard axial torch
Nebulizer typeSeaSpray
Nebulizer gas pressure220 kPa
Pump speed12 rpm
Sample uptake5 mL
Replicate read time30s
Number of replicates2
Sample delay time20s
Stabilization time15s
Rinse time40s
Background correctionOff-peak

Source: Agilent 710-ES Environmental Sample Analysis Application Note

All Agilent 710-ES documents by type

Questions answered in these manuals

What is the recommended pump rate for the VGA when operating with a sample solution?

A pump rate of 40 to 45 rpm is usually sufficient to ensure effective pumping and prevent liquid from rising in the gas-liquid separator.

From: Agilent 710-ES Determination of Mercury in Certified Reference Sludge Application Note

Which inert gas was used as the carrier gas for the VGA?

Argon was used as the VGA carrier gas, while nitrogen use is not recommended on ICP.

From: Agilent 710-ES Determination of Mercury in Certified Reference Sludge Application Note

What elements does the Agilent 710-ES analyze in this application?

The instrument analyzes 23 target analytes including mercury and cyanide, as well as elements like Ag, Al, As, Ba, Be, Ca, Cd, Co, Cr, Cu, Fe, K, Mg, Mn, Na, Ni, Pb, Sb, Se, Ti, V, and Zn.

From: Agilent 710-ES Environmental Sample Analysis Application Note

What is the maximum silver concentration recommended by the US EPA?

The US EPA recommends that the maximum silver concentration is limited to 2 mg/L.

From: Agilent 710-ES Environmental Sample Analysis Application Note

What is the maximum recommended heating temperature for decomposing plastic material?

The text states that a heating temperature of 95 °C is not high enough to decompose the plastic material.

From: Agilent 710-ES ICP-OES Application Note for Heavy Metal Monitoring

What digestion method is required to achieve complete recovery of mercury?

Closed vessel microwave digestion is the only method to get complete recovery of Hg.

From: Agilent 710-ES ICP-OES Application Note for Heavy Metal Monitoring