Agilent 7500ce ICP-MS Unmatched Spectral Interference Removal Using ORS with Helium Collision Mode Application Note

This document demonstrates the effective interference removal of a range of elements in a challenging and complex sample matrix using the Agilent 7500ce ICP-MS with a single set of He cell gas conditions, highlighting its high performance.


Agilent 7500ce ICP-MS Unmatched Spectral Interference Removal Using ORS with Helium Collision Mode Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
159 KB
Published
12 June, 2012
Updated
28 September, 2023
Model
7500CE
MD5 checksum
0028464F2D60E094C1F2CB1FC12368AF

About this manual

The document presents an application demonstrating that the Agilent 7500ce Inductively Coupled Plasma Mass Spectrometer (ICP-MS) can utilize a single set of Helium cell gas conditions to provide effective interference removal for various elements in complex sample matrices. It highlights the advantages of Kinetic Energy Discrimination (KED) over traditional reactive gas modes.

What's inside

  • Introduction
  • Helium Collision Mode
  • Advantages of He Collision Mode
  • Testing He Collision Mode
  • Instrumentation

Frequently asked questions

What is the primary function of the Agilent 7500ce ICP-MS when using Helium Collision Mode?

It removes all polyatomic species based on their size rather than their relative reactivity with a reaction gas.

How much Helium gas flow rate is required for the collision mode measurements?

5.5 mL/min of Helium gas only was added to the cell.

What specific condition must be met regarding ion energy for He collision mode to work properly?

The energy of all ions entering the cell must be very tightly controlled.