Agilent 7500ce ORS ICP-MS Carbonate Associated Sulfur Analysis Application Note

This report outlines preliminary methodology and results for the analysis of four samples for all three isotopes of sulfur in carbonate matrices using the 7500ce Octopole Reaction System (ORS) with xenon as the cell gas. Xenon was used to react with oxygen in order to attenuate the O2-derived polyatomic interferences at masses 32, 33, and 34. Consequently, each sulfur isotope was analyzed on-mass, leading to highly reproducible results.


Agilent 7500ce ORS ICP-MS Carbonate Associated Sulfur Analysis Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
148 KB
Published
02 July, 2012
Updated
30 September, 2023
Models
7500CE, 7500c
MD5 checksum
4DFBF216EB8EC4D32EA5DA9E45CE2C44

About this manual

This document outlines a methodology for determining carbonate associated sulfur (CAS) levels in seawater and pore water samples using the Agilent 7500ce Octopole Reaction System (ORS) ICP-MS. The study focuses on analyzing the three isotopes of sulfur (masses 32, 33, and 34) in carbonate matrices.

What's inside

  • Abstract
  • Introduction
  • Sulfur Analysis by ICP-MS
  • Sample Preparation and Instrumentation
  • Results and Discussion
  • Conclusions

Specifications

Forward power1500 W
Plasma gas flow15.0 L/min
Auxiliary gas flow1.0 L/min
Nebulizer gas flow0.80 L/min
Makeup gas flow0.20 L/min
Sampling depth8.0mm
Spray chamber temperature2°C
Xe cell gas flow0.25 mL/min

Frequently asked questions

How are polyatomic interferences removed for sulfur analysis?

Xenon is used as a reaction gas to react with oxygen, attenuating oxygen-derived polyatomic interferences at masses 32, 33, and 34.

What gas is used to remove polyatomic interferences during sulfur analysis?

Xenon is added as a cell gas to react with oxygen and attenuate interference species.

What are the detection limits for sulfur isotopes 32S and 34S?

The detection limit is 0.33 ppb for 32S and 1.7 ppb for 34S.

What is the sampling depth setting for the instrument?

The sampling depth is set to 8.0mm.