Agilent 720-ES 730-ES Series SmartRinse Maximizing Throughput Application Note Manual
SmartRinse is a feature within the Agilent ICP Expert II software designed to speed up sample wash-out in 700 Series ICP-OES instruments. It reduces carry-over and increases productivity by optimizing the rinse time between samples based on the specific element concentrations being measured, rather than using a fixed, worst-case time for all samples.
The feature works by measuring the baseline signal of selected element emission lines and calculating a threshold intensity. During the rinse phase, the system continuously monitors the signal intensity. Rinsing stops automatically once the signals for all selected elements fall below their calculated thresholds. The sensitivity of this process can be adjusted by the user via an SD factor setting to accommodate different analytical requirements.
Agilent 730-ES specifications
| Plasma gas flow | 15 L/min |
|---|---|
| Auxiliary gas flow | 1.5 L/min |
| Spraychamber type | Glass cyclonic |
| Torch type | Standard axial torch |
| Nebulizer type | SeaSpray |
| Sample uptake volume | 2.5 mL |
Source: Agilent 730-ES Environmental Sample Analysis Application Note
All Agilent 730-ES documents by type
How to use the product for one specific task, with worked examples.
Questions answered in these manuals
How does SmartRinse determine when to stop rinsing?
SmartRinse stops rinsing when the measured signal intensity of all nominated emission lines falls below a pre-determined threshold intensity calculated from baseline measurements.
From: Agilent 720-ES 730-ES Series SmartRinse Maximizing Throughput Application Note Manual
Can the sensitivity of SmartRinse be adjusted?
Yes, the sensitivity can be adjusted by changing the SD factor for each element in the ICP Expert II software to set a higher or lower threshold intensity.
From: Agilent 720-ES 730-ES Series SmartRinse Maximizing Throughput Application Note Manual
What is the maximum silver concentration recommended by US EPA?
The US EPA recommends that the maximum silver concentration be limited to 2 mg/L.
From: Agilent 730-ES Environmental Sample Analysis Application Note
What is the maximum concentration of silver recommended by the US EPA for this analysis?
The US EPA recommends that the maximum silver concentration be limited to 2 mg/L.
From: Agilent 730-ES Environmental Sample Analysis Application Note