Agilent 7500ce ICP-MS Chromium Speciation Application Note

This document describes a method for measuring chromium content. This method can separate and measure Cr(III) and Cr(VI), and provide an indication of the toxicity of Cr content in a sample.


Agilent 7500ce ICP-MS Chromium Speciation Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
8
File format
PDF
File size
497 KB
Published
12 June, 2012
Updated
30 September, 2023
Model
7500CE
MD5 checksum
26A2EF2816495547CA65CCF0029B94E7

About this manual

The document presents an optimized analytical method to determine the concentrations of trivalent chromium (Cr(III)) and hexavalent chromium (Cr(VI)) in natural samples. This is critical because Cr(VI) is toxic while Cr(III) is essential, requiring separate measurement for accurate toxicity assessment.

The proposed technique couples Ion Chromatography (IC) with Octopole Reaction Cell ICP-MS. The method utilizes an optimized sample stabilization protocol involving EDTA and incubation to maintain species integrity. It employs a specific IC column (Agilent part number G3268A) and mobile phase conditions to separate the Cr(III)-EDTA complex from Cr(VI). Detection is performed using the Agilent 7500ce ICP-MS in He cell gas mode to remove interferences, achieving detection limits below 20 ng/L.

What's inside

  • Introduction
  • Experimental
  • Chromatographic Conditions
  • ICP-MS Configuration
  • Results and Discussion
  • Detection Limits
  • Routine Analysis

Specifications

Column Part NumberG3268A
Mobile Phase Composition5 mM EDTA (2Na), pH 7
Flow Rate1.2 mL/min
Column TemperatureAmbient
Injection Volume Range50-500 uL
Reaction Temperature40 C
Incubation Time3h
EDTA Concentration in Mobile Phase5-15 mM

Frequently asked questions

What detection limits were achieved for Cr species?

Detection limits (DLs) of <20 ng/L were obtained for the individual Cr species.

How does the method handle interferences on mass 52?

The Agilent 7500ce ICP-MS uses He cell gas mode to remove ArC and C1OH interferences.

What is required for complete conversion of Cr(III) to Cr(III)-EDTA?

Complete conversion occurs after less than 1 hour at 60 C or 3 hours at 40 C.