Agilent XDB-C8 Manuals

ManualLib has 10 documents for the Agilent XDB-C8: 8 user manuals, 2 datasheets. Every file can be read online or downloaded as a PDF for free.

User Manual 8Datasheet 2

Key facts

Brand
Agilent
Model
XDB-C8
Industry
Industrial
Documents
10
Languages
EN
Last added
2012-07-04

Start here

Agilent ZORBAX Eclipse XDB-C8 RRHT Threaded Column Datasheet

Agilent ZORBAX Eclipse XDB-C8 RRHT Threaded Column Datasheet

Datasheet 2 pages · 89 KB · EN

The Agilent ZORBAX Eclipse XDB-C8 RRHT threaded columns are engineered for high-pressure operation up to 600 bar. They utilize eXtra-Dense Bonding (XDB) technology with dimethyl-n-octylsilane ligands bonded to ultra-pure Type B silica support to ensure high efficiency and stability. The packing is doubly endcapped to protect the silica surface from dissolution in mobile phases with intermediate and higher pH levels, making it suitable for separating acidic, basic, and highly polar compounds.

Read online

Agilent XDB-C8 specifications

Maximum Operating Pressure600 bar (9000 psi)
Maximum Operating Temperature60 °C
Optimum pH RangepH 3-8
Silica Support Surface Area180 m²/g
Controlled Pore Size80A

Source: Agilent ZORBAX Eclipse XDB-C8 RRHT Threaded Column Datasheet

All Agilent XDB-C8 documents by type

User Manual 8

Day-to-day operation, settings and troubleshooting. Start here if the product is already set up.

Questions answered in these manuals

What is the maximum operating pressure for these columns?

The maximum operating pressure is 600 bar (9000 psi).

From: Agilent ZORBAX Eclipse XDB-C8 RRHT Threaded Column Datasheet

What is the recommended pH range for optimum performance?

Optimum lifetime and performance are obtained at pH 3-8.

From: Agilent ZORBAX Eclipse XDB-C8 RRHT Threaded Column Datasheet

What is the recommended pH range for optimum performance?

Optimum lifetime and performance are obtained at a pH between 3 and 8.

From: Agilent ZORBAX Eclipse XDB-C8 Rapid Resolution Threaded Column Datasheet

Can the column be operated above 600 bar?

No, chromatographic performance has not been tested above this limit.

From: Agilent ZORBAX Eclipse XDB-C8 Rapid Resolution Threaded Column Datasheet

What is the recommended storage solvent for long-term storage?

Long-term storage should be in a pure organic solvent, preferably an aprotic liquid such as 100% acetonitrile.

From: Agilent ZORBAX Eclipse Eclipse XDB Eclipse XDB-C8 Rapid Resolution 3.5um 820116-006 600 bar Manual

Can the column be operated at temperatures up to 80 °C?

Yes, the material has been shown to be stable when operated at temperatures up to 80 °C.

From: Agilent ZORBAX Eclipse Eclipse XDB Eclipse XDB-C8 Rapid Resolution 3.5um 820116-006 600 bar Manual