Agilent HPLC Separation of Norethindrone Mestranol Using Progesterone as an Internal Standard on Eclipse XDB-CN (L10 USP) Eclipse XDB-C8 (L7 USP) Columns

This study investigated the comparison of USP-L10 and USP-L7 columns for dissolution and assay testing of norethindrone/mestranol tablets. The results showed that the USP-L10 column provided better peak separation for dissolution testing, while the USP-L7 column provided higher recovery for assay testing.


Agilent HPLC Separation of Norethindrone Mestranol Using Progesterone as an Internal Standard on Eclipse XDB-CN (L10 USP) Eclipse XDB-C8 (L7 USP) Columns - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
423 KB
Published
02 July, 2012
Updated
01 October, 2023
Model
XDB-C8
MD5 checksum
A01C9889F6EFC44117FB6A62631C0BBD

About this manual

This document presents an HPLC application for the separation of norethindrone and mestranol, utilizing progesterone as an internal standard. The study compares the performance of the Eclipse XDB-CN (L10) and Eclipse XDB-C8 (L7) columns against USP Formulary requirements for dissolution and assay testing of norethindrone/mestranol tablets.

The results indicate that both columns meet or exceed the specified performance measures, including theoretical plates, tailing factors, and resolution. The Eclipse XDB-CN column is highlighted for providing faster analysis times and lower costs while maintaining high-resolution separations suitable for both dissolution and assay methods.

What's inside

  • Introduction
  • Experimental
  • Results and Discussion
  • Highlights
  • USP Performance Measures
  • Conclusions

Specifications

Column Dimensions4.6 mm x 150 mm
Particle Size5 pm
Flow Rate1 mL/min
Temperature25 °C
Detection Wavelength205 nm
Injection Volume5 microliter

Frequently asked questions

What is used as an internal standard?

Progesterone is used as the internal standard (ISTD) for the assay method.

Does the Eclipse XDB-CN column meet USP dissolution requirements?

Yes, the Eclipse XDB-CN column easily meets the requirements for the dissolution method.