Agilent SepTech ST150 10-C18 Peptide Resolution Application Note

Agilent SepTech ST150 10-C18 Peptide Resolution Application Note

Application Note 3 pages · 721 KB · EN

Agilent Technologies investigated alternative solvents for reversed phase HPLC analysis of peptides due to global shortages in acetonitrile. This document evaluates ethanol as a viable substitute, comparing its performance against acetonitrile and three leading competitor columns.

The study utilizes the SepTech ST150 10-C18 column with a mixture of four peptides (Oxytocin, Angiotensin II, Angiotensin I, and Insulin). Results indicate that while ethanol provides similar solvent strength to acetonitrile, it alters selectivity, causing partial co-elution between insulin and angiotensin I. Additionally, ethanol exhibits higher UV absorbance at 220 nm compared to acetonitrile.

Read online

Agilent 10-C18 specifications

Column TypeSepTech ST150 10-C18
Flow Rate1 mL/min
Detection Wavelength (Acetonitrile)220 nm
Detection Wavelength (Methanol)280 nm

Source: Agilent Investigation into Alternatives to Acetonitrile for Peptide Analysis on ST150 10-C18 Application Note

All Agilent 10-C18 documents by type

Questions answered in these manuals

What is the stated particle size of the SepTech ST150 10-C18 column?

The column has a particle size of 10 micrometers.

From: Agilent SepTech ST150 10-C18 Peptide Resolution Application Note

What is the recommended flow rate for the peptide separation described?

The procedure specifies a flow rate of 1.0 mL/min.

From: Agilent SepTech ST150 10-C18 Peptide Resolution Application Note

Which alternative solvent is best for peptide analysis on the ST150 10-C18 column?

Methanol was found to be the best alternative, providing separation similar to acetonitrile with greater spacing between peptides.

From: Agilent Investigation into Alternatives to Acetonitrile for Peptide Analysis on ST150 10-C18 Application Note

Why must the detection wavelength be changed when using methanol?

Methanol has a high UV response at 220 nm, so the separation must be run at 280 nm to avoid background absorbance issues.

From: Agilent Investigation into Alternatives to Acetonitrile for Peptide Analysis on ST150 10-C18 Application Note