Agilent ZORBAX XDB-C8 Cephalosporin Separation Application Note

This document describes the separation of cephalosporins on Agilent ZORBAX XDB-C8 column.


Agilent ZORBAX XDB-C8 Cephalosporin Separation Application Note - cover page
Brand
Agilent
Category
Other
Document type
Application Note
Language
English
Pages
2
File format
PDF
File size
464 KB
Published
03 July, 2012
Updated
30 September, 2023
Model
XDB-C8
MD5 checksum
D48C4D070DF6FB44DC14DD4FA810EDCD

About this manual

This document describes a chromatographic separation method for five cephalosporins (Ceftazidime, Cefotaxime, Cefaclor, and two others) using a ZORBAX XDB-C8 column. The method operates at pH 7.0 to ensure good peak shape by avoiding partially ionized sample components.

The separation utilizes a mobile phase consisting of 85:15 25 mM Na2HPO4 (pH 7.0) and Acetonitrile (ACN). The analysis is performed with a 4.6x 150 mm column at 40°C, with a flow rate of 1 mL/min and UV detection at 254 nm.

What's inside

  • Application
  • Cephalosporins
  • ZORBAX XDB-C8
  • Chromatographic Separations
  • Mobile Phase
  • Operating Conditions

Specifications

Column Dimensions4.6x 150 mm
Mobile Phase Ratio85:15
Mobile Phase Concentration25 mM Na2HPO4
Injection Volume5pL
Flow Rate1 mL/min
Detection Wavelength254 nm

Frequently asked questions

What pH is recommended for separating cephalosporins on ZORBAX XDB-C8?

Separation is best achieved at pH 7.0 to avoid partially ionized sample components.

What is the composition of the mobile phase used in this application?

The mobile phase is a mixture of 85:15 25 mM Na2HPO4 (pH 7.0) and Acetonitrile (ACN).