Agilent 720-ES 730-ES Series SmartRinse Maximizing Throughput Application Note Manual

SmartRinse is a feature in the ICP Expert II software that speeds up sample wash-out, reduces carry-over, and increases productivity.


Agilent 720-ES 730-ES Series SmartRinse Maximizing Throughput Application Note Manual - cover page
Brand
Agilent
Category
Analyzers
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
111 KB
Published
05 July, 2012
Updated
28 September, 2023
Models
730-ES, 720-ES
MD5 checksum
3CAE6792C126A6F2A82D84C0BFA7270B

About this manual

SmartRinse is a feature within the Agilent ICP Expert II software designed to speed up sample wash-out in 700 Series ICP-OES instruments. It reduces carry-over and increases productivity by optimizing the rinse time between samples based on the specific element concentrations being measured, rather than using a fixed, worst-case time for all samples.

The feature works by measuring the baseline signal of selected element emission lines and calculating a threshold intensity. During the rinse phase, the system continuously monitors the signal intensity. Rinsing stops automatically once the signals for all selected elements fall below their calculated thresholds. The sensitivity of this process can be adjusted by the user via an SD factor setting to accommodate different analytical requirements.

What's inside

  • Technical Overview
  • How does SmartRinse work?
  • Adjusting the sensitivity of SmartRinse
  • Threshold Calculation
  • Wash-out Profiles

Specifications

Compatible AutosamplersAgilent SPS 3, CETAC ASX500 series and AIM1250

Frequently asked questions

How does SmartRinse determine when to stop rinsing?

SmartRinse stops rinsing when the measured signal intensity of all nominated emission lines falls below a pre-determined threshold intensity calculated from baseline measurements.

Can the sensitivity of SmartRinse be adjusted?

Yes, the sensitivity can be adjusted by changing the SD factor for each element in the ICP Expert II software to set a higher or lower threshold intensity.