Agilent 730-ES Environmental Sample Analysis Application Note

This application note describes the use of the Agilent 730-ES to carry out the US EPA/CLP compliant analysis of water samples. The instrument features true simultaneous measurement, allowing fast and accurate analysis of target analytes in water samples. It also has features such as extended linear range and long-term stability.


Agilent 730-ES Environmental Sample Analysis Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
10
File format
PDF
File size
165 KB
Published
12 June, 2012
Updated
28 September, 2023
Model
730-ES
MD5 checksum
5A59C468C75DF4B722E060122357D21C

About this manual

The document details a method using the Agilent 730-ES with a custom-designed CCD detector and an Agilent Switching Valve System (SVS) with SPS3 autosampler. The setup allows for simultaneous measurement of up to 1,000 elements from microgram per litre to percentage levels.

The note provides specific instrument operating conditions, including plasma gas flow at 15 L/min and auxiliary gas flow at 1.5 L/min. It also lists Method Detection Limits (MDL) and Contract Required Detection Limits (CRDL) for various elements such as Ag, Al, As, Ba, Be, Ca, Cd, Co, Cr, Cu, Fe, K, Mg, Mn, Na, Ni, Pb, Sb, Se, Tl, V, and Zn.

What's inside

  • Introduction
  • Instrumentation
  • Materials and Reagents
  • Standard and Sample Preparation
  • Instrument Operating Conditions
  • Method Detection Limits
  • Linear Range Analysis
  • Inter-Element Correction

Specifications

Plasma gas flow15 L/min
Auxiliary gas flow1.5 L/min
Spraychamber typeGlass cyclonic
Torch typeStandard axial torch
Nebulizer typeSeaSpray
Sample uptake volume2.5 mL

Frequently asked questions

What is the maximum silver concentration recommended by US EPA?

The US EPA recommends that the maximum silver concentration be limited to 2 mg/L.

What is the maximum concentration of silver recommended by the US EPA for this analysis?

The US EPA recommends that the maximum silver concentration be limited to 2 mg/L.

What are the required plasma gas and auxiliary gas flow rates for the instrument?

The operating conditions require a plasma gas flow of 15 L/min and an auxiliary gas flow of 1.5 L/min.

What is the sample uptake volume for the Agilent 730-ES in this application?

The instrument setting for sample uptake is 2.5 mL.