Agilent 710-ES Determination of Mercury in Certified Reference Sludge Application Note

This article describes the application of Agilent 710-ES in the determination of mercury in certified reference sludge materials. The method uses ICP-OES technology, which is simple to operate and has high accuracy. It also provides detailed operating steps and precautions for this method.


Agilent 710-ES Determination of Mercury in Certified Reference Sludge Application Note - cover page
Brand
Agilent
Document type
Application Note
Language
English
Pages
4
File format
PDF
File size
39 KB
Published
12 June, 2012
Updated
30 September, 2023
Model
710-ES
MD5 checksum
B8A3F4423C77AD8F69AEB78458F93AC8

About this manual

The document details the determination of mercury levels in solid wastes, specifically sewage sludges, which often contain high concentrations due to environmental contamination. It highlights the utility of vapor generation techniques coupled with Inductively Coupled Plasma-Optical Emission Spectrometry (ICP-OES) for measuring mercury at low concentrations typically found in environmental analysis.

The note presents a specific procedure using an Agilent 710-ES equipped with a megapixel CCD detector and an Agilent VGA-77P Vapor Generation Accessory. It outlines the operating conditions for both the ICP instrument and the VGA, including gas flows, power settings, and pump rates. The document also covers sample preparation methods involving acid digestion of NBS 2781 (Domestic Sludge) reference material and reports detection limits and analysis results.

What's inside

  • Introduction
  • Experimental
  • Standard Solutions and Reagents
  • Sample Preparation
  • Results and Discussion
  • Detection Limits
  • Analysis of a Certified Reference Material

Frequently asked questions

What is the recommended pump rate for the VGA when operating with a sample solution?

A pump rate of 40 to 45 rpm is usually sufficient to ensure effective pumping and prevent liquid from rising in the gas-liquid separator.

Which inert gas was used as the carrier gas for the VGA?

Argon was used as the VGA carrier gas, while nitrogen use is not recommended on ICP.

What are the specific flow rates for the plasma and auxiliary gases in this method?

The plasma gas flow was set to 13.5 L/min and the auxiliary gas flow was set to 1.5 L/min.

Download PDF (39 KB)

More Agilent Manuals


View all Agilent manuals →